Academic Company Events NI Developer Zone Support Solutions Products & Services Contact NI MyNI
KnowledgeBase

Category(s) Selected [start over]

Question Type
: Unexpected Behavior/ Bad Reading [remove] ,
Hardware
: SCXI [remove] ,
31-40 of 72
KnowledgeBase [see results below]. To refine your results, select another category.
Software
LabVIEW Modules (1)
LabWindows/CVI Development Systems (2)
Calibration Executive (1)
Driver Software (11)
Utility Software (2)
Hardware
Chassis (9)
Modules (38)
Terminal Blocks (14)
Document Languages
English (46)
Spanish (3)
Japanese (7)
Chinese (Simplified) (8)
Chinese (Traditional) (2)
Korean (6)

Sort by:
31-40 of 72
|
SCXI Configuration on Remote RT System Never Updates My SCXI chassis configuration in the Measurement & Automation Explorer (MAX) never changes even if modules are shuffled or changed. Why can't I delete my SCXI chassis from my RT system in MAX? URL: http://digital.ni.com/public.nsf/allkb/043D88990AB8535C86256FE80053ADA6
BNC-2096を使用してTEDSセンサの構成をするには、どのようにすればよいですか? TEDSセンサを持っているのですが、TEDSセンサとBNC-2096を使用するには、どのハードウェアが必要ですか?また、LabVIEWで集録するには、TEDSセンサの構成をどのように行ったらよいですか? URL: http://digital.ni.com/public.nsf/allkb/1C55B5F92D8D6CC186256FE200202F86
Erroneous or Noisy Thermocouple Readings Causing Hardware Trigger to Fail My thermocouple is returning noisy and erroneous readings when trying to trigger at a specified level using a hardware trigger within NI-DAQ and an SCXI-1102 module in an SCXI-1100 chassis. I am applying heat to the thermocouple by pinching it between my fingers. What can I do to eliminate URL: http://digital.ni.com/public.nsf/allkb/1000678F9D4B9E7F86256FD20059D053
Why Am I Getting Temperature Drift on My Thermocouples When My Cold Junction Temperature Changes? I am using a terminal block to connect my thermocouples to my data acquisition or signal conditioning device, but I notice that when the temperature of the cold junction (CJ) changes that my measurements drift. URL: http://digital.ni.com/public.nsf/allkb/C4920FDE8722A03D86256FCB0076DE6B
Why Does Measurement & Automation Explorer (MAX) Freeze When I Try to Auto-Detect My SCXI Modules Using Traditional DAQ? I go under Devices and Interfaces and right-click on Traditional DAQ Devices and Add an SCXI Chassis. I then follow the prompts selecting the Chassis Type, Chassis ID, and Chassis Address, but when I select Yes to Auto-Detect Modules Measurement & Automation Explorer (MAX) freezes. URL: http://digital.ni.com/public.nsf/allkb/0FA44B20777D77C186256F8700011D65
Thermocouple Reads a Large Negative Value When Connected to My SCXI Chassis Why does my thermocouple measure a value on the order of -1000 degrees when hooked up through my SCXI chassis? URL: http://digital.ni.com/public.nsf/allkb/F84C4728FF407E0886256F1800568241
Potential Analog Input Offset Using an Externally Powered Pressure Transducer in an SCXI System I am using an SCXI system for analog input (for example an SCXI-1100 with an SCXI-1300 terminal block) and measuring voltage from a pressure transducer that requires an external power source. Why do I see a constant offset over the entire range of my measurement? URL: http://digital.ni.com/public.nsf/allkb/4BB6C879B6055E7386256ED300676D72
Why Am I Getting # inf or # j When Running A Shunt Calibration With My SCXI 1520 Module? I have a SCXI 1520 strain module with a 1314 terminal block. I am certain that my strain gauge is wired in correctly but I receive a # inf or # j error when running a shunt calibration. My strain gauge is a quarter bridge I bridge type. URL: http://digital.ni.com/public.nsf/allkb/33F09F8A218914B386256E9800518842
What Are The Pinouts of the SCXI-1180? NI Home 2 ratings: 4 out of 5 Rate this Document What Are The Pinouts of the SCXI-1180? Hardware: SCXI>>Terminal Blocks Problem: The SCXI-1180 manual does not list the pinouts for the module. Solution: The SCXI-1180 pinouts can be found in the following table: 50-pin Connector E-Series MIO Board URL: http://digital.ni.com/public.nsf/allkb/B2889ADCFD4B915986256E830081790F
Triggering with PFI 7 on the SCXI-1600 What is PFI 7 on the SCXI-1600? Why does my computer crash when I attempt to trigger off it? URL: http://digital.ni.com/public.nsf/allkb/95EFEE3EF3E643EE86256E82006BAF18
Sort by:
31-40 of 72
|