Last Revised: 2010-12-22 10:36:50.0
|
|
|
|
OS Information
|
Driver Information
|
Software Compatibility
|
The USB-9219 uses six-position spring-terminal connectors in each channel for direct signal connectivity. You can purchase additional connectors to reduce signal connection time for multiple test units. In addition to extra connectors, a strain relief kit is available to secure the signal wires.
You can use the NI 9219 module included in the USB-9219 kit in a CompactRIO chassis or an NI CompactDAQ USB chassis for higher-channel-count systems.
Strain relief backshells for signal wire security and high-voltage protection (qty 4): NI 9972
Extra connectors for six-position connector modules (qty 10): NI 9973
For information on special configurations and pricing, visit ni.com/oem.
LabVIEW Professional Development System for Windows
|
|
NI LabWindows™/CVI for Windows
|
|
NI Measurement Studio Professional Edition
|
|
NI system assurance programs are designed to make it even easier for you to own an NI system. These programs include configuration and deployment services for your NI PXI, CompactRIO, or Compact FieldPoint system. The NI Basic System Assurance Program provides a simple integration test and ensures that your system is delivered completely assembled in one box. When you configure your system with the NI Standard System Assurance Program, you can select from available NI system driver sets and application development environments to create customized, reorderable software configurations. Your system arrives fully assembled and tested in one box with your software preinstalled. When you order your system with the standard program, you also receive system-specific documentation including a bill of materials, an integration test report, a recommended maintenance plan, and frequently asked question documents. Finally, the standard program reduces the total cost of owning an NI system by providing three years of warranty coverage and calibration service. Use the online product advisors at ni.com/advisor to find a system assurance program to meet your needs.
NI measurement hardware is calibrated to ensure measurement accuracy and verify that the device meets its published specifications. To ensure the ongoing accuracy of your measurement hardware, NI offers basic or detailed recalibration service that provides ongoing ISO 9001 audit compliance and confidence in your measurements. To learn more about NI calibration services or to locate a qualified service center near you, contact your local sales office or visit ni.com/calibration.
Get answers to your technical questions using the following National Instruments resources.
While you may never need your hardware repaired, NI understands that unexpected events may lead to necessary repairs. NI offers repair services performed by highly trained technicians who quickly return your device with the guarantee that it will perform to factory specifications. For more information, visit ni.com/repair.
The NI training and certification program delivers the fastest, most certain route to increased proficiency and productivity using NI software and hardware. Training builds the skills to more efficiently develop robust, maintainable applications, while certification validates your knowledge and ability.
NI offers options for extending the standard product warranty to meet the life-cycle requirements of your project. In addition, because NI understands that your requirements may change, the extended warranty is flexible in length and easily renewed. For more information, visit ni.com/warranty.
NI offers design-in consulting and product integration assistance if you need NI products for OEM applications. For information about special pricing and services for OEM customers, visit ni.com/oem.
Our Professional Services Team is comprised of NI applications engineers, NI Consulting Services, and a worldwide National Instruments Alliance Partner program of more than 700 independent consultants and integrators. Services range from start-up assistance to turnkey system integration. Visit ni.com/alliance.
The following specifications are typical for the range 0 to 60 °C unless otherwise noted.
| Input Characteristics | ||
|
Number of channels |
4 analog input channels | |
|
ADC resolution |
24 bits | |
|
Type of ADC |
Delta-sigma (with analog prefiltering) | |
|
Sampling mode |
Simultaneous | |
|
Type of TEDS supported |
IEEE 1451.4 TEDS Class II (Interface) | |
| Mode input ranges | ||
|---|---|---|
| Mode | Nominal Range(s) | Actual Range(s) |
| Voltage | ±60 V, ±15 V, ±4 V, ±1 V, ±125 mV | ±60 V, ±15 V, ±4 V, ±1 V, ±125 mV |
| Current | ±25 mA | ±25 mA |
| 4-Wire and 2-Wire Resistance | 10 kΩ, 1 kΩ | 10.5 kΩ, 1.05 kΩ |
| Thermocouple | ±125 mV | ±125 mV |
| 4-Wire and 3-Wire RTD | Pt 1,000, Pt 100 | 5.05 kΩ, 505 Ω |
| Quarter-Bridge | 350 Ω, 120 Ω | 390 Ω, 150 Ω |
| Half-Bridge | ±500 mV/V | ±500 mV/V |
| Full-Bridge | ±62.5 mV/V, ±7.8 mV/V | ±62.5 mV/V, ±7.8125 mV/V |
|
Conversion time, no channels in TC mode |
||
|
High Speed |
10 ms for all channels | |
|
Best 60 Hz Rejection |
110 ms for all channels | |
|
Best 50 Hz Rejection |
130 ms for all channels | |
|
High Resolution |
500 ms for all channels | |
|
Conversion time, one or more channels in TC mode |
||
|
High Speed |
20 ms for all channels | |
|
Best 60 Hz Rejection |
120 ms for all channels | |
|
Best 50 Hz Rejection |
140 ms for all channels | |
|
High Resolution |
510 ms for all channels | |
|
Overvoltage protection |
||
|
Terminals 1 and 2 |
±30 V | |
|
Terminals 3 through 6, across any combination |
±60 V | |
|
Input impedance |
||
|
Voltage and Digital In modes (±60 V, ±15 V, ±4 V) |
1 MΩ | |
|
Current mode |
<40 Ω | |
|
All other modes |
>1 GΩ | |
| Accuracy | ||
|---|---|---|
| Mode, Range | Gain Error (% of Reading) 2 | Offset Error (ppm of Range) 2 |
| Voltage, ±60 V | ±0.3, ±0.4 | ±20, ±50 |
| Voltage, ±15 V | ±0.3, ±0.4 | ±60, ±180 |
| Voltage, ±4 V | ±0.3, ±0.4 | ±240, ±720 |
| Voltage, ±1 V | ±0.1, ±0.18 | ±15, ±45 |
| Voltage/Thermocouple, ±125 mV | ±0.1, ±0.18 | ±120, ±360 |
| Current, ±25 mA | ±0.1, ±0.6 | ±30, ±100 |
| 4-Wire and 2-Wire Resistance, 10 kΩ 1 | ±0.1, ±0.5 | ±120, ±320 |
| 4-Wire and 2-Wire Resistance, 1 kΩ 1 | ±0.1, ±0.5 | ±1,200, ±3,200 |
| 4-Wire and 3-Wire RTD, Pt 1,000 | ±0.1, ±0.5 | ±240, ±640 |
| 4-Wire and 3-Wire RTD, Pt 100 | ±0.1, ±0.5 | ±2,400, ±6,400 |
| Quarter-Bridge, 350 Ω | ±0.1, ±0.5 | ±2,400, ±6,400 |
| Quarter-Bridge, 120 Ω | ±0.1, ±0.5 | ±2,400, ±6,400 |
| Half-Bridge, ±500 mV/V | ±0.03, ±0.07 | ±300, ±450 |
| Full-Bridge, ±62.5 mV/V | ±0.03, ±0.08 | ±300, ±1,000 |
| Full-Bridge, ±7.8 mV/V | ±0.03, ±0.08 | ±2,200, ±8,000 |
|
Cold-junction compensation sensor accuracy |
±1°C typ |
| Stability | ||
|---|---|---|
| Mode, Range | Gain Drift (ppm of Reading/°C) | Offset Drift (ppm of Range/°C) |
| Voltage, ±60 V | ±20 | ±0.2 |
| Voltage, ±15 V | ±20 | ±0.8 |
| Voltage, ±4 V | ±20 | ±3.2 |
| Voltage, ±1 V | ±10 | ±0.2 |
| Voltage/Thermocouple, ±125 mV | ±10 | ±1.6 |
| Current, ±25 mA | ±15 | ±0.4 |
| 4-Wire and 2-Wire Resistance, 10 kΩ | ±15 | ±3 |
| 4-Wire and 2-Wire Resistance, 1 kΩ | ±15 | ±30 |
| 4-Wire and 3-Wire RTD, Pt 1,000 | ±15 | ±6 |
| 4-Wire and 3-Wire RTD, Pt 100 | ±15 | ±60 |
| Quarter-Bridge, 350 Ω | ±15 | ±120 |
| Quarter-Bridge, 120 Ω | ±15 | ±240 |
| Half-Bridge, ±500 mV/V | ±3 | ±20 |
| Full-Bridge, ±62.5 mV/V | ±3 | ±20 |
| Full-Bridge, ±7.8 mV/V | ±3 | ±20 |
| Input noise in ppm of Rangerms | ||||
|---|---|---|---|---|
| Mode, Range | Conversion Time | |||
| High Speed | Best 60 Hz Rejection | Best 50 Hz Rejection | High Resolution | |
| Voltage, ±60 V | 7.6 | 1.3 | 1.3 | 0.5 |
| Voltage, ±15 V | 10.8 | 1.9 | 1.9 | 0.7 |
| Voltage, ±4 V | 10.8 | 2.7 | 2.7 | 1.3 |
| Voltage, ±1 V | 7.6 | 1.3 | 1.3 | 0.5 |
| Voltage/Thermocouple, 125 mV | 10.8 | 1.9 | 1.9 | 1.0 |
| Current, ±25 mA | 10.8 | 1.9 | 1.9 | 1.0 |
| 4-Wire and 2-Wire Resistance, 10 kΩ | 4.1 | 1.3 | 0.8 | 0.3 |
| 4-Wire and 2-Wire Resistance, 1 kΩ | 7.1 | 1.8 | 1.2 | 0.7 |
| 4-Wire and 3-Wire RTD, Pt 1,000 | 7.6 | 1.7 | 1.1 | 0.4 |
| 4-Wire and 3-Wire RTD, Pt 100 | 10.8 | 1.9 | 1.9 | 0.9 |
| Quarter-Bridge, 350 Ω | 5.4 | 1.0 | 1.0 | 0.7 |
| Quarter-Bridge, 120 Ω | 5.4 | 1.0 | 1.0 | 0.7 |
| Half-Bridge, ±500 mV/V | 3.8 | 0.5 | 0.5 | 0.2 |
| Full-Bridge, ±62.5 mV/V | 5.4 | 1.0 | 1.0 | 0.8 |
| Full-Bridge, ±7.8 mV/V | 30 | 4.7 | 4.7 | 2.3 |
|
Input bias current |
<1 nA | |
|
INL |
±15 ppm | |
|
CMRR (ƒin = 60 Hz) |
>100 dB | |
|
NMRR |
||
|
Best 60 Hz Rejection |
90 dB at 60 Hz | |
|
Best 50 Hz Rejection |
80 dB at 50 Hz | |
|
High Resolution |
65 dB at 50 Hz and 60 Hz | |
| Excitation level for Half-Bridge and Full-Bridge modes | ||
|---|---|---|
| Mode | Load Resistance (Ω) | Excitation (V) |
| Half-Bridge | 700 | 2.5 |
| Half-Bridge | 240 | 2.0 |
| Full-Bridge | 350 | 2.7 |
| Full-Bridge | 120 | 2.2 |
| Excitation level for Resistance, RTD, and Quarter-Bridge modes | |
|---|---|
| Load Resistance (Ω) | Excitation (mV) |
| 120 | 50 |
| 350 | 150 |
| 1,000 | 430 |
| 10,000 | 2,200 |
| Power Requirements | ||
|
Current consumption from USB |
500 mA, max | |
|
Suspend mode |
2.5 MA, max | |
| Bus Interface | ||
|
USB specification |
USB 2.0 Hi-Speed | |
| Physical Characteristics | ||
|
Spring-terminal wiring |
18 to 28 AWG copper conductor wire with 7 mm (0.28 in.) of insulation stripped from the end | |
|
Weight |
258 g (9.1 oz) | |
| Safety | ||
| If you need to clean the module, wipe it with a dry towel. | ||
| Safety Voltages | ||
| Connect only voltages that are within these limits. | ||
|
Isolation |
||
|
Channel-to-channel |
||
|
Continuous |
250 VAC, Measurement Category II | |
|
Withstand |
1390 VAC, verified by a 5 s dielectric withstand test | |
|
Channel-to-earth ground |
||
|
Continuous |
250 VAC, Measurement Category II | |
|
Withstand |
2300 VAC, verified by a 5 s dielectric withstand test | |
| Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet (e.g., 115 V for U.S. or 230 V for Europe). Examples of Measurement Category II are measurements performed on household appliances, portable tools, and similar products. | ||
| Safety Standards | ||
|
This product is designed to meet the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use:
Note For UL and other safety certifications, refer to the product label or the Online Product Certification section. |
||
| Hazardous Locations | ||
| This device is not certified for use in hazardous locations. | ||
| Environmental | ||
| The USB-9219 is intended for indoor use only but may be used outdoors if installed in a suitable enclosure. | ||
|
Operating temperature (IEC 60068-2-1, IEC 60068-2-2) |
0 to 60 °C | |
|
Storage temperature (IEC 60068-2-1, IEC 60068-2-2) |
–40 to 85 °C | |
|
Ingress protection |
IP 40 | |
|
Operating humidity (IEC 60068-2-56) |
10 to 90% RH, noncondensing | |
|
Storage humidity (IEC 60068-2-56) |
5 to 95% RH, noncondensing | |
|
Maximum altitude |
2,000 m | |
|
Pollution Degree (IEC 60664) |
2 | |
| Electromagnetic Compatibility | ||
|
This product is designed to meet the requirements of the following standards of EMC for electrical equipment for measurement, control, and laboratory use:
Note For EMC compliance, operate this device with shielded cables. |
||
| CE Compliance | ||
|
This product meets the essential requirements of applicable European Directives, as amended for CE marking, as follows:
Note For the standards applied to assess the EMC of this product, refer to the Online Product Certification section. |
||
| Online Product Certification | ||
| Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by module number or product line, and click the appropriate link in the Certification column. | ||
| Environmental Management | ||
|
National Instruments is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial not only to the environment but also to NI customers. For additional environmental information, refer to the NI and the Environment Web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document. |
||
| Waste Electrical and Electronic Equipment (WEEE) | ||
|
EU Customers At the end of their life cycle, all products must be sent to a WEEE recycling center. For more information about WEEE recycling centers and National Instruments WEEE initiatives, visit ni.com/environment/weee.htm. ![]() |
||
| Calibration | ||
| You can obtain the calibration certificate for this device at ni.com/calibration. | ||
|
Calibration interval |
1 year | |
![]() |
|||||||||||||||||||||||||||
| NI USB 9219 Module | |||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||
©2010 National Instruments. All rights reserved. CompactRIO, FieldPoint, LabVIEW, National Instruments, National Instruments Alliance Partner, NI, ni.com, and NI-DAQ are trademarks of National Instruments. Other product and company names listed are trademarks or trade names of their respective companies. A National Instruments Alliance Partner is a business entity independent from National Instruments and has no agency, partnership, or joint-venture relationship with National Instruments.