Last Revised: 2009-07-29 14:12:31.0
|
|
|
|
OS Information
|
Driver Information
|
Software Compatibility
|
| Family | Bus | Analog Inputs | AI Resolution (Bits) | Analog Outputs | AO Resolution (Bits) | Max AO Update Rate (MS/s) | AO Range | Digital I/O | Correlated (clocked) DIO |
|---|---|---|---|---|---|---|---|---|---|
| NI 6280 | PCI, PXI | 16 | 18 | - | - | - | - | 24 | 8, up to 10 MHz |
| NI 6281 | PCI, PXI | 16 | 18 | 2 | 16 | 2.8 | Programmable per channel | 24 | 8, up to 10 MHz |
| NI 6284 | PCI, PXI | 32 | 18 | - | - | - | - | 48 | 32, up to 10 MHz |
| NI 6289 | PCI, PXI | 32 | 18 | 4 | 16 | 2.8 | Programmable per channel | 48 | 32, up to 10 MHz |
For test, you can use M Series high-accuracy analog inputs and 10 MHz digital lines with NI signal conditioning for applications including electronics test, component characterization, and sensor and signal measurements requiring instrument-class accuracy. The 18-bit analog-to-digital converter (ADC) and available filtering provide a 4 times increase in resolution and 5 times more measurement sensitivity. With fast sampling rates and a low noise floor, these devices can accurately acquire dynamic signals. For better noise rejection, the onboard lowpass filters significantly improve device accuracy. Advanced analog clamping circuitry protects the hardware from overvoltage conditions and ensures accurate measurements on nonsaturated channels. High-accuracy M Series devices are compatible with NI SCC and SCXI signal conditioning platforms, which provide amplification, filtering, and power for virtually every type of sensor. These platforms also are compliant with IEEE 1451.4 smart transducer electronic data sheet (TEDS) sensors, which offer digital storage for sensor data sheet information.
M Series digital lines can drive 24 mA for relay and actuator control. By clocking the digital lines as fast as 10 MHz, you can use these lines for pulse-width modulation (PWM) to control valves, motors, fans, lamps, and pumps. With four waveform analog outputs, two 80 MHz counter/timers, and six DMA channels, M Series devices can execute multiple control loops simultaneously. The analog outputs on the high-accuracy M Series devices can generate up to 2.86 MS/s and provide user-defined programmable offsets and ranges for maximum waveform resolution over any custom range. High-accuracy M Series devices also have direct support for encoder measurements, protected digital lines, and digital debounce filters for control applications. With up to 32 analog inputs, 32 clocked digital lines, and four analog outputs, you can execute multiple control loops with a single device. For higher-count control loops, you can use M Series devices in conjunction and tightly synchronized with National Instruments analog output devices for 64 or more loops. With the NI SoftMotion Development Module for LabVIEW, you can create a complete custom motion controller with M Series devices.
For design applications, you can use the wide range of I/O - from 32 analog inputs to 48 digital lines - to measure and verify prototype designs. M Series devices and NI LabVIEW SignalExpress interactive measurement software deliver benchtop measurements to the PC. With LabVIEW SignalExpress interactive configuration-based steps, you can quickly create design verification tests. The fast acquisition and generation rates of high-speed M Series devices along with LabVIEW SignalExpress provide on-the-fly design analysis. You can convert your tested and verified LabVIEW SignalExpress projects to LabVIEW applications for immediate M Series DAQ use and bridge the gap between test, control, and design applications.
M Series modules for PXI are hybrid-slot-compatible so that you can use them in both PXI slots and the hybrid slots found in new PXI Express chassis. The PXI Systems Alliance specifies that hybrid-slot-compatible PXI modules use modified slot connectors to mechanically fit in both PXI slots and hybrid slots. This mechanical change:
However, hybrid-slot-compatible PXI modules do not include the pins used to implement PXI local bus communication, which is used for backplane SCXI control from the right-most PXI slot in PXI/SCXI combination chassis (NI PXI-1010, PXI-1011, PXI-1050, and PXI-1052). For these applications, NI provides unmodified M Series PXI modules that maintain the required local bus capabilities. Refer to the SCXI Control of PXI/SCXI Combination Chassis section in the Ordering Information section for part numbers.
When you need to obtain performance from a data acquisition device beyond the capabilities of a multifunction DAQ device, National Instruments provides simultaneous sampling with the S Series and intelligent DAQ with the R Series. The S Series architecture dedicates an ADC per channel to provide higher aggregate sampling rates compared to multiplexed devices. S Series devices are ideal for applications including IF digitization, transient recording, ultrasound and sonar testing, and high-energy physics.
R Series multifunction DAQ devices contain a 1M/3M gate field-programmable gate array (FPGA) that is reconfigurable using the NI LabVIEW FPGA Module. These devices have up to eight independent 16-bit analog inputs with up to 200 kHz simultaneous sampling, up to eight independent 16-bit analog outputs with up to 1 MHz simultaneous update rates, and up to 96 digital I/O lines configurable at rates up to 40 MHz. You can customize these devices to develop capabilities such as complete control over the synchronization and timing of all signals and operations; user-defined onboard decision-making logic; and digital lines individually configurable as input, output, counter/timers, PWM, flexible encoder inputs, or user-defined communication protocols.
Signal conditioning is required for sensor measurements or voltage inputs greater than 10 V. NI SCXI is a versatile, high-performance signal conditioning platform, optimized for high-channel-count applications. NI SCC provides portable, flexible signal conditioning options on a per-channel basis. Visit ni.com/sigcon for resources on available NI signal conditioning.
For a complete list of accessories, visit the product page on ni.com.
| Products | Part Number | Recommended Accessories | Part Number |
| NI PCI-6280 | |||
| NI PCI-6280 Each NI PCI-6280 requires: 1 Cable, 1 Connector Block | 779108-01 | Cable: Shielded - SHC68-68-EPM Cable (2m) **Also available: Unshielded | 192061-02 |
| Connector Block: Screw Terminals - SCB-68 **Also available: BNC Termination | 776844-01 | ||
| NI PXI-6280 | |||
| NI PXI-6280 Each NI PXI-6280 requires: 1 Cable, 1 Connector Block | 779120-01 | Cable: Shielded - SHC68-68-EPM Cable (2m) **Also available: Unshielded | 192061-02 |
| Connector Block: Screw Terminals - SCB-68 **Also available: BNC Termination | 776844-01 | ||
| NI PCI-6281 | |||
| NI PCI-6281 Each NI PCI-6281 requires: 1 Cable, 1 Connector Block | 779109-01 | Cable: Shielded - SHC68-68-EPM Cable (2m) **Also available: Unshielded | 192061-02 |
| Connector Block: Screw Terminals - SCB-68 **Also available: BNC Termination | 776844-01 | ||
| NI PXI-6281 | |||
| NI PXI-6281 Each NI PXI-6281 requires: 1 Cable, 1 Connector Block | 779121-01 | Cable: Shielded - SHC68-68-EPM Cable (2m) **Also available: Unshielded | 192061-02 |
| Connector Block: Screw Terminals - SCB-68 **Also available: BNC Termination | 776844-01 | ||
| NI PCI-6284 | |||
| NI PCI-6284 Each NI PCI-6284 requires: 2 Cables, 2 Connector Blocks | 779110-01 | Connector 0: | |
| Cable: Shielded - SHC68-68-EPM Cable (2m) **Also available: Unshielded | 192061-02 | ||
| Connector Block: Screw Terminals - SCB-68 **Also available: BNC Termination | 776844-01 | ||
| Connector 1: | |||
| Cable: Shielded - SHC68-68-EPM Cable (2m) **Also available: Unshielded | 192061-02 | ||
| Connector Block: Screw Terminals - SCB-68 **Also available: BNC Termination | 776844-01 | ||
| NI PXI-6284 | |||
| NI PXI-6284 Each NI PXI-6284 requires: 2 Cables, 2 Connector Blocks | 779122-01 | Connector 0: | |
| Cable: Shielded - SHC68-68-EPM Cable (2m) **Also available: Unshielded | 192061-02 | ||
| Connector Block: Screw Terminals - SCB-68 **Also available: BNC Termination | 776844-01 | ||
| Connector 1: | |||
| Cable: Shielded - SHC68-68-EPM Cable (2m) **Also available: Unshielded | 192061-02 | ||
| Connector Block: Screw Terminals - SCB-68 **Also available: BNC Termination | 776844-01 | ||
| NI PCI-6289 | |||
| NI PCI-6289 Each NI PCI-6289 requires: 2 Cables, 2 Connector Blocks | 779111-01 | Connector 0: | |
| Cable: Shielded - SHC68-68-EPM Cable (2m) **Also available: Unshielded | 192061-02 | ||
| Connector Block: Screw Terminals - SCB-68 **Also available: BNC Termination | 776844-01 | ||
| Connector 1: | |||
| Cable: Shielded - SHC68-68-EPM Cable (2m) **Also available: Unshielded | 192061-02 | ||
| Connector Block: Screw Terminals - SCB-68 **Also available: BNC Termination | 776844-01 | ||
| NI PXI-6289 | |||
| NI PXI-6289 Each NI PXI-6289 requires: 2 Cables, 2 Connector Blocks | 779639-01 | Connector 0: | |
| Cable: Shielded - SHC68-68-EPM Cable (2m) **Also available: Unshielded | 192061-02 | ||
| Connector Block: Screw Terminals - SCB-68 **Also available: BNC Termination | 776844-01 | ||
| Connector 1: | |||
| Cable: Shielded - SHC68-68-EPM Cable (2m) **Also available: Unshielded | 192061-02 | ||
| Connector Block: Screw Terminals - SCB-68 **Also available: BNC Termination | 776844-01 | ||
| NI PXI-6289 for SCXI Control in PXI/SCXI Combination Chassis Each NI PXI-6289 for SCXI Control in PXI/SCXI Combination Chassis requires: 2 Cables, 2 Connector Blocks | 779123-01 | Connector 0: | |
| Cable: Shielded - SHC68-68-EPM Cable (2m) **Also available: Unshielded | 192061-02 | ||
| Connector Block: Screw Terminals - SCB-68 **Also available: BNC Termination | 776844-01 | ||
| Connector 1: | |||
| Cable: Shielded - SHC68-68-EPM Cable (2m) **Also available: Unshielded | 192061-02 | ||
| Connector Block: Screw Terminals - SCB-68 **Also available: BNC Termination | 776844-01 | ||
NI LabVIEW Professional Development System for Windows
|
|
NI LabWindows™/CVI for Windows
|
|
NI LabVIEW SignalExpress for Windows
|
|
NI Measurement Studio Professional Edition
|
|
NI system assurance programs are designed to make it even easier for you to own an NI system. These programs include configuration and deployment services for your NI PXI, CompactRIO, or Compact FieldPoint system. The NI Basic System Assurance Program provides a simple integration test and ensures that your system is delivered completely assembled in one box. When you configure your system with the NI Standard System Assurance Program, you can select from available NI system driver sets and application development environments to create customized, reorderable software configurations. Your system arrives fully assembled and tested in one box with your software preinstalled. When you order your system with the standard program, you also receive system-specific documentation including a bill of materials, an integration test report, a recommended maintenance plan, and frequently asked question documents. Finally, the standard program reduces the total cost of owning an NI system by providing three years of warranty coverage and calibration service. Use the online product advisors at ni.com/advisor to find a system assurance program to meet your needs.
NI measurement hardware is calibrated to ensure measurement accuracy and verify that the device meets its published specifications. NI offers a number of calibration services to help maintain the ongoing accuracy of your measurement hardware. These services allow you to be completely confident in your measurements, and help you maintain compliance to standards like ISO 9001, ANSI/NCSL Z540-1 and ISO/IEC 17025. To learn more about NI calibration services or to locate a qualified service center near you, contact your local sales office or visit ni.com/calibration.
Get answers to your technical questions using the following National Instruments resources.
While you may never need your hardware repaired, NI understands that unexpected events may lead to necessary repairs. NI offers repair services performed by highly trained technicians who quickly return your device with the guarantee that it will perform to factory specifications. For more information, visit ni.com/repair.
The NI training and certification program delivers the fastest, most certain route to increased proficiency and productivity using NI software and hardware. Training builds the skills to more efficiently develop robust, maintainable applications, while certification validates your knowledge and ability.
NI offers options for extending the standard product warranty to meet the life-cycle requirements of your project. In addition, because NI understands that your requirements may change, the extended warranty is flexible in length and easily renewed. For more information, visit ni.com/warranty.
NI offers design-in consulting and product integration assistance if you need NI products for OEM applications. For information about special pricing and services for OEM customers, visit ni.com/oem.
Our Professional Services Team is comprised of NI applications engineers, NI Consulting Services, and a worldwide National Instruments Alliance Partner program of more than 600 independent consultants and integrators. Services range from start-up assistance to turnkey system integration. Visit ni.com/alliance.
| Analog Input | ||
|
Number of channels |
|
|
|
NI 6280/6281 |
|
8 differential or 16 single ended |
|
NI 6284/6289 |
|
16 differential or 32 single ended |
|
ADC resolution |
|
18 bits |
|
DNL |
|
No missing codes guaranteed |
|
INL |
|
Refer to the AI Absolute Accuracy Tables |
|
Sampling rate |
|
|
|
Maximum |
|
625 kS/s single channel, |
|
Minimum |
|
No minimum |
|
Timing accuracy |
|
50 ppm of sample rate |
|
Timing resolution |
|
50 ns |
|
Input coupling |
|
DC |
|
Input range |
±10 V, ±5 V, ± 2 V, ±1 V, ±0.5 V, ± 0.2 V, ± 0.1 V |
|
|
Maximum working voltage for analog inputs |
±11 V of AI GND |
|
|
CMRR (DC to 60 Hz) |
110 dB |
|
|
Input impedance |
||
|
Device on |
|
|
|
AI+ to AI GND |
>10 GΩ in parallel with 100 pF |
|
|
AI- to AI GND |
>10 GΩ in parallel with 100 pF |
|
|
Device off |
||
|
AI+ to AI GND |
820 Ω |
|
|
AI- to AI GND |
820 Ω |
|
|
Input bias current |
±100 pA |
|
|
Crosstalk (at 100 kHz) |
|
|
|
Adjacent channels |
|
-75 dB |
|
Non-adjacent channels |
|
-95 dB |
|
Small signal bandwidth (-3 dB) |
|
750 kHz filter off, |
|
Input FIFO size |
|
2,047 samples |
|
Scan list memory |
|
4,095 entries |
|
Data transfers |
|
|
|
PCI/PXI devices |
|
DMA (scatter-gather), interrupts, programmed I/O |
|
USB devices |
|
USB Signal Stream, programmed I/O |
|
Overvoltage protection (AI <0..31>, AI SENSE, AI SENSE 2) |
|
|
|
Device on |
|
±25 V for up to eight AI pins |
|
Device off |
|
±15 V for up to eight AI pins |
|
Input current during overvoltage condition |
|
±20 mA max/AI pin |
|
Settling Time for Multichannel Measurements |
||
| Range |
Filter Off |
Filter On |
|
|---|---|---|---|
|
±15 ppm of Step |
±4 ppm of Step |
±4 ppm of Step |
|
| ±10 V, ±5 V | 2 μs | 8 μs | 50 μs |
| ±2 V, ±1 V, ±0.5 V | 2.5 μs | 8 μs | 50 μs |
| ±0.2 V, ±0.1 V | 3 μs | 8 μs | 50 μs |
| Typical Performance Graphs | |
![]() |
![]() |
![]() |
|
|
Analog Triggers |
||
|
Number of triggers |
|
1 |
|
Source |
|
|
|
NI 6280/6281 |
|
AI <0..15>, APFI 0 |
|
NI 6284/6289 |
|
AI <0..31>, APFI <0..1> |
|
Functions |
|
Start Trigger, Reference Trigger, Pause Trigger, Sample Clock, Convert Clock, Sample Clock Timebase |
|
Source level |
|
|
|
AI <0..31> |
|
±Full scale |
|
APFI <0..1> |
|
±10 V |
|
Resolution |
|
10 bits, 1 in 1024 |
|
Modes |
|
Analog edge triggering, analog edge triggering with hysteresis, and analog window triggering |
|
Bandwidth (-3 dB) |
|
|
|
AI <0..31> |
|
700 kHz filter off, |
|
APFI <0..1> |
|
5 MHz |
|
Accuracy |
|
±1% |
|
APFI <0..1> characteristics |
|
|
|
Input impedance |
|
10 kΩ |
|
Coupling |
|
DC |
|
Protection |
|
|
|
Power on |
|
±30 V |
|
Power off |
|
±15 V |
|
Analog Output |
||
|
Number of channels |
|
|
|
NI 6280/6284 |
|
0 |
|
NI 6281 |
|
2 |
|
NI 6289 |
|
4 |
|
DAC resolution |
16 bits | |
|
DNL |
|
±1 LSB |
|
Monotonicity |
|
16 bit guaranteed |
|
Accuracy |
|
Refer to the AO Absolute Accuracy Table |
|
Maximum update rate |
|
|
|
1 channel |
|
2.86 MS/s |
|
2 channels |
|
2.00 MS/s |
|
3 channels |
|
1.54 MS/s |
|
4 channels |
|
1.25 MS/s |
|
Timing accuracy |
|
50 ppm of sample rate |
|
Timing resolution |
|
50 ns |
|
Output range |
Offset ± reference, includes ±10 V, ±5 V, ±2 V, and ±1 V calibrated ranges | |
|
Offset |
|
0 V, 5 V, APFI <0..1>, AO <0..3>1 |
|
Reference |
|
10 V, 5 V, 2 V, 1 V, APFI <0..1>, AO <0..3>1 |
|
Maximum output level |
|
±11 V |
|
Output coupling |
DC |
|
|
Output impedance |
0.2 Ω |
|
|
Output current drive |
±5 mA | |
|
Overdrive protection |
±25 V |
|
|
Overdrive current |
20 mA |
|
|
Power-on state |
±5 mV2 |
|
|
Power-on glitch |
2.3 V peak for 1.2 s |
|
|
Output FIFO size |
8,191 samples shared among channels used |
|
|
Data transfers |
|
|
|
PCI/PXI devices |
|
DMA (scatter-gather), interrupts, programmed I/O |
|
USB devices |
|
USB Signal Stream, programmed I/O |
|
AO waveform modes:
|
||
|
Settling time, full scale step 15 ppm (1 LSB) |
3 μs |
|
|
Slew rate |
20 V/μs |
|
|
Glitch energy at midscale transition, ±10 V range |
|
|
|
Magnitude |
|
15 mV |
|
Duration |
|
0.5 μs |
|
1 An AO channel cannot be a reference or offset to itself. 2 For all USB-6281/6289 Screw Terminal devices, when powered on, the analog output signal is not defined until after USB |
||
|
External Reference |
||
|
APFI <0..1> characteristics |
|
|
|
Input impedance |
|
10 kΩ |
|
Coupling |
|
DC |
|
Protection |
|
|
|
Power on |
|
±30 V |
|
Power off |
|
±15 V |
|
Range |
|
±11 V |
|
|
||
|
Calibration (AI and AO) |
||
|
Recommended warm-up time |
|
|
|
PCI/PXI devices |
|
15 minutes |
|
USB devices |
|
30 minutes |
|
Calibration interval |
|
2 years |
AI Absolute Accuracy Table (Filter On) |
||||||||||
| Nominal Range | Residual Gain Error (ppm of Reading) |
Gain Tempco (ppm/°C) |
Reference Tempco |
Residual Offset Error (ppm of Range) |
Offset Tempco (ppm of Range/°C) |
INL Error (ppm of Range) |
Random Noise, σ (µVrms) |
Absolute Accuracy at Full Scale1 (µV) |
Sensitivity2 (µV) |
|
|---|---|---|---|---|---|---|---|---|---|---|
|
Positive |
Negative |
|||||||||
| 10 | -10 | 40 | 17 | 1 | 8 | 11 | 10 | 60 | 980 | 24 |
| 5 | -5 | 45 | 17 | 1 | 8 | 11 | 10 | 30 | 510 | 12 |
| 2 | -2 | 45 | 17 | 1 | 8 | 13 | 10 | 12 | 210 | 4.8 |
| 1 | -1 | 55 | 17 | 1 | 15 | 15 | 10 | 7 | 120 | 2.8 |
| 0.5 | -0.5 | 55 | 17 | 1 | 30 | 20 | 10 | 4 | 70 | 1.6 |
| 0.2 | -0.2 | 75 | 17 | 1 | 45 | 35 | 10 | 3 | 39 | 1.2 |
| 0.1 | -0.1 | 120 | 17 | 1 | 60 | 60 | 10 | 2 | 28 | 0.8 |
|
Accuracies listed are valid for up to two years from the device external calibration. AbsoluteAccuracy = Reading · (GainError) + Range · (OffsetError) + NoiseUncertainty GainError = ResidualAIGainError + GainTempco · (TempChangeFromLastInternalCal) + ReferenceTempco · (TempChangeFromLastExternalCal) OffsetError = ResidualAIOffsetError + OffsetTempco · (TempChangeFromLastInternalCal) + INL_Error
1 Absolute accuracy at full scale on the analog input channels is determined using the following assumptions: TempChangeFromLastExternalCal = 10 °C TempChangeFromLastInternalCal = 1 °C number_of_readings = 100 CoverageFactor = 3 σ For example, on the 10 V range, the absolute accuracy at full scale is as follows: GainError = 40 ppm + 17 ppm · 1 + 1 ppm · 10 GainError = 67 ppm OffsetError = 8 ppm + 11 ppm · 1 + 10 ppm OffsetError = 29 ppm
AbsoluteAccuracy = 10 V · (GainError) + 10 V · (OffsetError) + NoiseUncertainty AbsoluteAccuracy = 980 μV 2 Sensitivity is the smallest voltage change that can be detected. It is a function of noise. |
||||||||||
AI Absolute Accuracy Table (Filter Off) |
||||||||||
| Nominal Range | Residual Gain Error (ppm of Reading) |
Gain Tempco (ppm/°C) |
Reference Tempco |
Residual Offset Error (ppm of Range) |
Offset Tempco (ppm of Range/°C) |
INL Error (ppm of Range) |
Random Noise, σ (µVrms) |
Absolute Accuracy at Full Scale1 (µV) |
Sensitivity2 (µV) |
|
|---|---|---|---|---|---|---|---|---|---|---|
|
Positive |
Negative |
|||||||||
| 10 | -10 | 45 | 17 | 1 | 10 | 11 | 10 | 70 | 1050 | 28.0 |
| 5 | -5 | 50 | 17 | 1 | 10 | 11 | 10 | 35 | 550 | 14.0 |
| 2 | -2 | 50 | 17 | 1 | 10 | 13 | 10 | 15 | 230 | 6.0 |
| 1 | -1 | 60 | 17 | 1 | 17 | 15 | 10 | 12 | 130 | 4.8 |
| 0.5 | -0.5 | 60 | 17 | 1 | 32 | 20 | 10 | 10 | 80 | 4.0 |
| 0.2 | -0.2 | 80 | 17 | 1 | 47 | 35 | 10 | 9 | 43 | 3.6 |
| 0.1 | -0.1 | 120 | 17 | 1 | 62 | 60 | 10 | 9 | 31 | 3.6 |
|
Accuracies listed are valid for up to two years from the device external calibration. AbsoluteAccuracy = Reading · (GainError) + Range · (OffsetError) + NoiseUncertainty GainError = ResidualAIGainError + GainTempco · (TempChangeFromLastInternalCal) + ReferenceTempco · (TempChangeFromLastExternalCal) OffsetError = ResidualAIOffsetError + OffsetTempco · (TempChangeFromLastInternalCal) + INL_Error
1 Absolute accuracy at full scale on the analog input channels is determined using the following assumptions: TempChangeFromLastExternalCal = 10 °C TempChangeFromLastInternalCal = 1 °C number_of_readings = 100 CoverageFactor = 3 σ For example, on the 10 V range, the absolute accuracy at full scale is as follows: GainError = 45 ppm + 17 ppm · 1 + 1 ppm · 10 GainError = 72 ppm OffsetError = 10 ppm + 11 ppm · 1 + 10 ppm OffsetError = 31 ppm
AbsoluteAccuracy = 10 V · (GainError) + 10 V · (OffsetError) + NoiseUncertainty AbsoluteAccuracy = 1050 μV 2 Sensitivity is the smallest voltage change that can be detected. It is a function of noise. |
||||||||||
AO Absolute Accuracy Table |
||||||||
| Nominal Range | Residual Gain Error (ppm of Reading) |
Gain Tempco (ppm/°C) |
Reference Tempco |
Residual Offset Error (ppm of Range) |
Offset Tempco (ppm of Range/°C) |
INL Error (ppm of Range) |
Absolute Accuracy at Full Scale1 (µV) |
|
|---|---|---|---|---|---|---|---|---|
|
Positive |
Negative |
|||||||
| 10 | -10 | 55 | 15 | 1 | 30 | 12 | 32 | 1540 |
| 5 | -5 | 60 | 15 | 1 | 30 | 17 | 32 | 820 |
| 2 | -2 | 65 | 25 | 1 | 40 | 30 | 32 | 404 |
| 1 | -1 | 85 | 25 | 1 | 57 | 50 | 32 | 259 |
|
1 Absolute Accuracy at full scale numbers is valid immediately following internal calibration and assumes the device is operating within 10 °C of the last external calibration. AbsoluteAccuracy = OutputValue · (GainError) + Range · (OffsetError) GainError = ResidualGainError + GainTempco · (TempChangeFromLastInternalCal) + ReferenceTempco · (TempChangeFromLastExternalCal) OffsetError = ResidualOffsetError + AOOffsetTempco · (TempChangeFromLastInternalCal) + INL_Error |
||||||||
|
Digital I/O/PFI |
||
|
Static Characteristics |
||
|
Number of channels |
|
|
|
NI 6280/6281 |
|
24 total |
|
NI 6284/6289 |
48 total |
|
|
I/O type |
5 V TTL/CMOS compatible |
|
|
Ground reference |
D GND |
|
|
Direction control |
|
Each terminal individually programmable as input or output |
|
Pull-down resistor |
|
50 kΩ typical, 20 kΩ minimum |
|
Input voltage protection3 |
|
±20 V on up to two pins |
| 3 Stresses beyond those listed under Input voltage protection may cause permanent damage to the device. | ||
|
Waveform Characteristics (Port 0 Only) |
||
|
Terminals used |
|
|
|
NI 6280/6281 |
|
Port 0 (P0.<0..7>) |
|
NI 6284/6289 |
|
Port 0 (P0.<0..31>) |
|
Port/sample size |
|
|
|
NI 6280/6281 |
|
Up to 8 bits |
|
NI 6284/6289 |
|
Up to 32 bits |
|
Waveform generation (DO) FIFO |
|
2,047 samples |
|
Waveform acquisition (DI) FIFO |
|
2,047 samples |
|
DI Sample Clock frequency |
|
|
|
PCI/PXI devices |
|
0 to 10 MHz4 |
|
USB devices |
|
0 to 1 MHz system dependent4 |
|
DO Sample Clock frequency |
|
|
|
PCI/PXI devices |
|
|
|
Regenerate from FIFO |
|
0 to 10 MHz |
|
Streaming from memory |
|
0 to 10 MHz system dependent4 |
|
USB devices |
|
|
|
Regenerate from FIFO |
|
0 to 10 MHz |
|
Streaming from memory |
|
0 to 1 MHz system dependent4 |
|
Data transfers |
|
|
|
PCI/PXI devices |
|
DMA (scatter-gather), interrupts, programmed I/O |
|
USB devices |
|
USB Signal Stream, programmed I/O |
|
DO or DI Sample Clock source5 |
|
Any PFI, RTSI, AI Sample or Convert Clock, AO Sample Clock, Ctr n Internal Output, and many other signals |
| 4 Performance can be dependent on bus latency and volume of bus activity. 5 The digital subsystem does not have its own dedicated internal timing engine. Therefore, a sample clock must be provided from another subsystem on the device or an external source. |
||
|
PFI/Port 1/Port 2 Functionality |
||
|
Functionality |
|
Static digital input, static digital output, timing input, timing output |
|
Timing output sources |
|
Many AI, AO, counter, DI, DO timing signals |
|
Debounce filter settings |
|
125 ns, 6.425 µs, 2.56 ms, disable; high and low transitions; selectable per input |
|
Recommended Operation Conditions |
||
|
Level |
Min |
Max |
|---|---|---|
| Input high voltage (VIH) | 2.2 V | 5.25 V |
| Input low voltage (VIL) | 0 V | 0.8 V |
| Output high current (IOH) P0.<0..31> PFI <0..15>/P1/P2 |
— — |
-24 mA -16 mA |
| Output low current (IOL) P0.<0..31> PFI <0..15>/P1/P2 |
— — |
24 mA 16 mA |
|
Electrical Characteristics |
||
|
Level |
Min |
Max |
|---|---|---|
| Positive-going threshold (VT+) Negative-going threshold (VT-) |
— 0.8 V |
2.2 V — |
| Delta VT hysteresis (VT+ - VT-) | 0.2 V | — |
|
IIL input low current (Vin = 0 V) IIH input high current (Vin = 5 V) |
— — |
-10 µA 250 µA |
|
Digital I/O Characteristics |
|
|
|
|
![]() |
![]() |
|
General-Purpose Counter/Timers |
||
|
Number of counter/timers |
|
2 |
|
Resolution |
|
32 bits |
|
Counter measurements |
|
Edge counting, pulse, semi-period, period, two-edge separation |
|
Position measurements |
|
X1, X2, X4 quadrature encoding with Channel Z reloading; two-pulse encoding |
|
Output applications |
|
Pulse, pulse train with dynamic updates, frequency division, equivalent time sampling |
|
Internal base clocks |
|
80 MHz, 20 MHz, 0.1 MHz |
|
External base clock frequency |
|
0 MHz to 20 MHz |
|
Base clock accuracy |
|
50 ppm |
|
Inputs |
|
Gate, Source, HW_Arm, Aux, A, B, Z, Up_Down |
|
Routing options for inputs |
|
Any PFI, RTSI, PXI_TRIG, PXI_STAR, analog trigger, many internal signals |
|
FIFO |
|
2 samples |
|
Data transfers |
|
|
|
PCI/PCIe/PXI/PXIe devices |
|
Dedicated scatter-gather DMA controller for each counter/timer; interrupts, programmed I/O |
|
USB devices |
|
USB Signal Stream, programmed I/O |
|
Frequency Generator |
||
|
Number of channels |
|
1 |
|
Base clocks |
|
10 MHz, 100 kHz |
|
Divisors |
|
1 to 16 |
|
Base clock accuracy |
50 ppm |
|
|
Output can be available on any PFI or RTSI terminal. |
||
|
Phase-Locked Loop (PLL) (PCI/PXI Devices Only) |
||
|
Number of PLLs |
|
1 |
|
Reference signal |
|
PXI_STAR, PXI_CLK10, RTSI <0..7> |
|
Output of PLL |
|
80 MHz Timebase; other signals derived from 80 MHz Timebase including 20 MHz and 100 kHz Timebases |
|
External Digital Triggers |
||
|
Source |
|
Any PFI, RTSI, PXI_TRIG, PXI_STAR |
|
Polarity |
|
Software-selectable for most signals |
|
Analog input function |
|
Start Trigger, Reference Trigger, Pause Trigger, Sample Clock, Convert Clock, Sample Clock Timebase |
|
Analog output function |
|
Start Trigger, Pause Trigger, Sample Clock, Sample Clock Timebase |
|
Counter/timer functions |
|
Gate, Source, HW_Arm, Aux, A, B, Z, Up_Down |
|
Digital waveform generation (DO) function |
Sample Clock |
|
|
Digital waveform acquisition (DI) function |
Sample Clock |
|
|
Device-To-Device Trigger Bus |
||
|
PCI devices |
|
RTSI <0..7>6 |
|
PXI devices |
|
PXI_TRIG <0..7>, PXI_STAR |
|
USB devices |
|
None |
|
Output selections |
10 MHz Clock; frequency generator output; many internal signals |
|
|
Debounce filter settings |
125 ns, 6.425 µs, 2.56 ms, disable; high and low transitions; selectable per input |
|
|
6 In other sections of this document, RTSI refers to RTSI <0..7> for PCI devices or PXI_TRIG <0..7> for PXI devices. |
||
|
Bus Interface |
||
|
PCI/PXI devices |
|
3.3 V or 5 V signal environment |
|
USB devices |
|
USB 2.0 Hi-Speed or full-speed7,8 |
|
DMA channels (PCI/PXI devices) |
|
6, analog input, analog output, digital input, digital output, counter/timer 0, counter/timer 1 |
|
USB Signal Stream (USB devices) |
4, can be used for analog input, analog output, digital input, digital output, counter/timer 0, counter/timer 1 |
|
|
All PXI-628x devices support one of the following features:
|
||
|
Table 1. PXI/SCXI Combo and PXI Express Chassis Compatibility |
||
|
M Series Device |
M Series Part Number |
SCXI Control in PXI/SCXI Combo Chassis |
PXI Express Hybrid Slot Compatible |
|---|---|---|---|
| PXI-6280 | 191501C-04 | No | Yes |
| PXI-6281 | 191501C-03 | No | Yes |
| PXI-6284 | 191501C-02 | No | Yes |
| PXI-6289 | 191501C-01 | No | Yes |
| 191501C-11 | Yes | No | |
| Earlier versions of PXI-628x | 191501A-0x 191501B-0x |
Yes | No |
|
7 If you are using a USB M Series device in full-speed mode, device performance will be lower and you will not be able to achieve maximum sampling/update rates. |
||
|
Power Requirements |
||
|
PCI/PXI devices |
|
|
|
Current draw from bus during no-load condition9 |
|
|
|
+5 V |
|
0.03 A |
|
+3.3 V |
|
0.78 A |
|
+12 V |
|
0.40 A |
|
-12 V |
|
0.06 A |
|
PCI/PXI devices |
|
|
|
Current draw from bus during AI and AO overvoltage condition9 |
|
|
|
+5 V |
|
0.03 A |
|
+3.3 V |
|
1.26 A |
|
+12 V |
|
0.43 A |
|
-12 V |
|
0.06 A |
Caution USB-628x devices must be powered with NI offered AC adapter or a National Electric Code (NEC) Class evice and has appropriate safety certification marks for country of use.
|
USB devices |
|
|
|
Power supply requirements |
|
11 to 30 VDC, 20 W, locking or non-locking power jack with 0.080 in. diameter center pin, 5/16-32 thread for locking collars |
|
Power supply fuse |
|
2 A, 250 V |
|
9 Does not include P0/PFI/P1/P2 and +5 V terminals. |
||
|
Power Limits |
||
|
Caution Exceeding the power limits may cause unpredictable behavior by the device and/or PC/chassis. |
||
|
PCI devices |
|
|
|
+5 V terminal (connector 0) |
|
1 A max10 |
|
+5 V terminal (connector 1) |
|
1 A max10 |
|
PXI devices |
|
|
|
+5 V terminal (connector 0) |
1 A max10 |
|
|
+5 V terminal (connector 1) |
|
1 A max10 |
|
P0/PFI/P1/P2 and +5 V terminals combined |
|
2 A max |
|
USB devices |
|
|
|
+5 V terminal |
|
1 A max11 |
|
P0/PFI/P1/P2 and +5 V terminals combined |
|
2 A max |
|
10 Older revisions have a self-resetting fuse that opens when current exceeds this specification. Newer revisions have a traditional fuse that opens when current exceeds this specification. This fuse is not customer-replaceable; if the fuse permanently opens, return the device to NI for repair. |
||
|
Physical Requirements |
||
|
Printed circuit board dimensions |
|
|
|
NI PCI-6280/6281/6284/6289 |
|
10.6 cm x 15.5 cm (4.2 in. x 6.1 in.) |
|
NI PXI-6280/6281/6284/6289 |
|
Standard 3U PXI |
|
Enclosure dimensions (includes connectors) |
|
|
|
NI USB-6281/6289 Mass Termination |
|
18.8 x 17.09 x 4.45 cm (7.4 x 6.73 x 1.75 in.) |
|
NI USB-6281/6289 Screw Terminal |
|
26.67 x 17.09 x 4.45 cm (10.5 x 6.73 x 1.75 in.) |
|
NI USB-6258/6289 OEM |
|
Refer to the NI USB-622x/625x/628x OEM User Guide |
|
Weight |
|
|
|
NI PCI-6280 |
|
151 g (5.3 oz) |
|
NI PCI-6281 |
|
158 g (5.6 oz) |
|
NI PCI-6284 |
|
159 g (5.6 oz) |
|
NI PCI-6289 |
|
167 g (5.9 oz) |
|
NI PXI-6280 |
|
218 g (7.7 oz) |
|
NI PXI-6281 |
|
225 g (7.9 oz) |
|
NI PXI-6284 |
|
229 g (8.1 oz) |
|
NI PXI-6289 |
|
237 g (8.4 oz) |
|
NI USB-6281 Mass Termination |
|
1.04 kg (2 lb 4.5 oz) |
|
NI USB-6289 Mass Termination |
|
1.06 kg (2 lb 5.5 oz) |
|
NI USB-6281 OEM |
|
261 g (9.2 oz) |
|
NI USB-6289 OEM |
|
274 g (9.6 oz) |
|
NI USB-6281 Screw Terminal |
|
1.46 kg (3 lb 3.4 oz) |
|
NI USB-6289 Screw Terminal |
|
1.52 kg (3 lb 5.5 oz) |
|
I/O connector |
|
|
|
NI PCI/PXI-6280/6281 |
|
1 68-pin VHDCI |
|
NI PCI/PXI-6284/6289 |
|
2 68-pin VHDCI |
|
NI USB-6281 Mass Termination |
|
1 68-pin SCSI |
|
NI USB-6289 Mass Termination |
|
2 68-pin SCSI |
|
NI USB-6281 OEM |
|
1 34-pin IDC, 1 50-pin IDC |
|
NI USB-6289 OEM |
|
2 34-pin IDC, 2 50-pin IDC |
|
NI USB-6281 Screw Terminal |
|
64 screw terminals |
|
NI USB-6289 Screw Terminal |
|
128 screw terminals |
|
USB-6281/6289 Screw Terminal wiring |
16-28 AWG | |
Maximum Working Voltage12 |
||
|
NI 6280/6281/6284/6289 channel-to-earth |
|
11 V, Measurement Category I |
|
Caution Do not use for measurements within Categories II, III, or IV. |
||
|
12 Maximum working voltage refers to the signal voltage plus the common-mode voltage. |
||
Environmental |
||
|
Operating temperature |
|
|
|
PCI/PXI devices |
|
0 to 55 °C |
|
USB devices |
|
0 to 45 °C |
|
Storage temperature |
|
-20 to 70 °C |
|
Humidity |
|
10 to 90% RH, noncondensing |
|
Maximum altitude |
|
2,000 m |
|
Pollution Degree (indoor use only) |
|
2 |
|
Shock and Vibration (PXI Devices Only) |
||
|
Operational shock |
|
30 g peak, half-sine, 11 ms pulse |
|
Random vibration |
|
|
|
Operating |
|
5 to 500 Hz, 0.3 grms |
|
Nonoperating |
|
5 to 500 Hz, 2.4 grms |
Safety |
||
|
This product is designed to meet the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use: |
||
|
||
|
Note For UL and other safety certifications, refer to the product label or the Online Product Certification section |
||
|
Electromagnetic Compatibility |
||
|
This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use: |
||
|
||
|
Note For the standards applied to assess the EMC of this product, refer to the Online Product Certification section. |
||
|
Note For EMC compliance, operate this product according to the documentation. |
||
|
Note For EMC compliance, operate this device with shielded cables. |
||
|
CE Compliance |
||
|
This product meets the essential requirements of applicable European Directives as follows: |
||
|
||
| Online Product Certification | ||
|
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by model number or product line, and click the appropriate link in the Certification column. |
||
| Environmental Management | ||
|
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial not only to the environment but also to NI customers. For additional environmental information, refer to the NI and the Environment Web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document. |
||
| Waste Electrical and Electronic Equipment (WEEE) | ||
|
EU Customers At the end of their life cycle, all products must be sent to a WEEE recycling center. For more information about WEEE recycling centers and National Instruments WEEE initiatives, visit ni.com/environment/weee.htm. |
||
![]() |
![]() |
| NI PCI/PXI-6280 Pinout |
![]() |
| NI PCI/PXI-6281 Pinout |
![]() |
| NI PCI/PXI-6284 Pinout |
![]() |
| NI PCI/PXI-6289 Pinout |
©2008 National Instruments. All rights reserved. CVI, DIAdem, LabVIEW, Measurement Studio, National Instruments, National Instruments Alliance Partner, NI, ni.com, NI SoftMotion, SCXI, and SignalExpress are trademarks of National Instruments. The mark LabWindows is used under a license from Microsoft Corporation. Windows is a registered trademark of Microsoft Corporation in the United States and other countries. Linux® is the registered trademark of Linus Torvalds in the U.S. and other countries. Other product and company names listed are trademarks or trade names of their respective companies. A National Instruments Alliance Partner is a business entity independent from National Instruments and has no agency, partnership, or joint-venture relationship with National Instruments.