150 MHz, 200 MS/s, 12-Bit Digitizers

NI PXI-5124, NI PCI-5124

  • 2 channels simultaneously sampled with 12-bit resolution
  • 200 MS/s real-time sample rate and 4.0 GS/s random interleaved sampling
  • 150 MHz bandwidth
  • 200 mVpp to 20 Vpp input ranges
  • >75 dBc SFDR
  • 8, 32, 256, or 512 MB of memory per channel
  • Edge, window, hysteresis, video, and digital triggering with 50 ps timestamping
 

Overview

NI 5124 high-resolution digitizers feature two 200 MS/s simultaneously sampled input channels with 12-bit resolution, 150 MHz bandwidth, and up to 512 MB of memory per channel in a 3U PXI or PCI device. NI 5124 devices use the high-speed PCI bus and the scatter-gather bus mastering of the NI MITE ASIC to move data to the computer at speeds up to 100 times faster than traditional instrument interfaces, thereby dramatically decreasing overall test time. With the NI Synchronization and Memory Core (SMC) architecture of an NI 5124, you can create mixed-signal systems using signal generators and digital waveform generators/analyzers or build a high-channel-count digitizer system with subnanosecond synchronization between channels.

Back to Top
 

Application and Technology

My First Composite Component

Dual 200 MS/s, 12-Bit Input Channels for Time and Frequency Analysis
• 150 MHz input bandwidth with antialias and noise filters
• >75 dBc spurious-free dynamic range (SFDR)
• 4.0 GS/s equivalent time sampling for repetitive signals
• Independent channel-selectable 200 mVpp to 20 Vpp input ranges
• Independent channel-selectable 50 Ω or 1 MΩ input impedance
• 2-year calibration cycle and 0 to 55 °C operating temperature

Deep Onboard Memory
• 8, 32, 256, or 512 MB of memory per channel
• Capture more than 1 million triggered waveforms with multiple-record hardware rearm
• Stream data continuously from onboard memory to host memory or disk

Triggering, Clocking, and Synchronization
• Edge, window, hysteresis, video, digital, triggering with 50 ps timestamping
• Pretrigger and posttrigger acquisition in single and multiple-record mode
• Internal 200 MHz clock or external clock from 50 to 210 MHz
• Phase lock to PXI 10 MHz reference or external reference from 1 to 20 MHz


Back to Top
 

Support and Services

System Assurance Programs

NI system assurance programs are designed to make it even easier for you to own an NI system. These programs include configuration and deployment services for your NI PXI, CompactRIO, or Compact FieldPoint system. The NI Basic System Assurance Program provides a simple integration test and ensures that your system is delivered completely assembled in one box. When you configure your system with the NI Standard System Assurance Program, you can select from available NI system driver sets and application development environments to create customized, reorderable software configurations. Your system arrives fully assembled and tested in one box with your software preinstalled. When you order your system with the standard program, you also receive system-specific documentation including a bill of materials, an integration test report, a recommended maintenance plan, and frequently asked question documents. Finally, the standard program reduces the total cost of owning an NI system by providing three years of warranty coverage and calibration service. Use the online product advisors at ni.com/advisor to find a system assurance program to meet your needs.



Calibration

NI measurement hardware is calibrated to ensure measurement accuracy and verify that the device meets its published specifications. To ensure the ongoing accuracy of your measurement hardware, NI offers basic or detailed recalibration service that provides ongoing ISO 9001 audit compliance and confidence in your measurements. To learn more about NI calibration services or to locate a qualified service center near you, contact your local sales office or visit ni.com/calibration.


Technical Support

Get answers to your technical questions using the following National Instruments resources.

  • Support - Visit ni.com/support to access the NI KnowledgeBase, example programs, and tutorials or to contact our applications engineers who are located in NI sales offices around the world and speak the local language.
  • Discussion Forums - Visit forums.ni.com for a diverse set of discussion boards on topics you care about.
  • Online Community - Visit community.ni.com to find, contribute, or collaborate on customer-contributed technical content with users like you.


Repair

While you may never need your hardware repaired, NI understands that unexpected events may lead to necessary repairs. NI offers repair services performed by highly trained technicians who quickly return your device with the guarantee that it will perform to factory specifications. For more information, visit ni.com/repair.


Training and Certifications

The NI training and certification program delivers the fastest, most certain route to increased proficiency and productivity using NI software and hardware. Training builds the skills to more efficiently develop robust, maintainable applications, while certification validates your knowledge and ability.

  • Classroom training in cities worldwide - the most comprehensive hands-on training taught by engineers.
  • On-site training at your facility - an excellent option to train multiple employees at the same time.
  • Online instructor-led training - lower-cost, remote training if classroom or on-site courses are not possible.
  • Course kits - lowest-cost, self-paced training that you can use as reference guides.
  • Training memberships and training credits - to buy now and schedule training later.
Visit ni.com/training for more information.



Extended Warranty

NI offers options for extending the standard product warranty to meet the life-cycle requirements of your project. In addition, because NI understands that your requirements may change, the extended warranty is flexible in length and easily renewed. For more information, visit ni.com/warranty.


OEM

NI offers design-in consulting and product integration assistance if you need NI products for OEM applications. For information about special pricing and services for OEM customers, visit ni.com/oem.


Alliance

Our Professional Services Team is comprised of NI applications engineers, NI Consulting Services, and a worldwide National Instruments Alliance Partner program of more than 700 independent consultants and integrators. Services range from start-up assistance to turnkey system integration. Visit ni.com/alliance.


Back to Top
 

Detailed Specifications

12-Bit 200 MS/s Digitizer

This topic lists the specifications for the NI PXI/PCI-5124 (NI 5124) high-speed digitizer. Unless otherwise noted, these specifications are valid for the following conditions:

  • All filter settings
  • All impedance selections
  • Sample clock set to 200 MS/s using onboard clock

Typical values are representative of an average unit operating at room temperature. Specifications are subject to change without notice. For the most recent NI 5124 specifications, visit ni.com/manuals.

To access the NI 5124 documentation, including the NI High-Speed Digitizers Getting Started Guide, which contains functional descriptions of the NI 5124 signals, navigate to Start»All Programs»National Instruments»NI-SCOPE»Documentation.

Hot Surface If the NI 5124 has been in use, it may exceed safe handling temperatures and cause burns. Allow the NI 5124 to cool before removing it from the PXI chassis or PC. Refer to the Environment section for operating temperatures of this device.

Vertical
Analog Input (Channel 0 and Channel 1)
Specification Value Comments
Number of Channels Two (simultaneously sampled)
Connector BNC
Impedance and Coupling
Input Impedance 50 Ω  ±2.0% 1 MΩ ± 0.75% in parallel with a typical capacitance of 29 pF Software selectable
Input Coupling AC, DC, GND AC coupling available on 1 MΩ only
Voltage Levels
Full Scale (FS) Input Range and Programmable Vertical Offset 50 Ω 1 MΩ
Range
(Vpk-pk)
Vertical
Offset Range
(V)
Range
(Vpk-pk)
Vertical
Offset Range
(V)
0.2 ±0.1 0.2 ±0.1
0.4 ±0.2 0.4 ±0.2
1 ±0.5 1 ±0.5
2 ±1 2 ±1
4 ±2 4 ±2
10 0 10 ±5
20 0
Maximum Input Overload 50 Ω 1 MΩ
7 Vrms with |Peaks| 10 V |Peaks| 42 V
Accuracy
Resolution 12 bits
DC Accuracy (Programmable Vertical Offset = 0 V) Range (Vpk-pk) NI PXI-5124 NI PCI-5124 Within ±5 °C of self-calibration temperature
0.2 and 0.4 ±(0.65% of Input + 1.3 mV) ±(0.65% of Input + 1.8 mV)
1 and 2 ±(0.65% of Input + 1.5 mV) ±(0.65% of Input + 2.1 mV)
4, 10, and 20 (1 MΩ only) ±(0.65% of Input + 10.0 mV) ±(0.65% of Input + 10.0 mV)
Programmable Vertical Offset Accuracy ±0.4% of offset setting Within ±5 °C of self-calibration temperature
DC Drift Range (Vpk-pk) 50 Ω and 1 MΩ
0.2, 0.4, 1, and 2 ±(0.057% of Input + 0.006% of FS + 100 μV) per °C
4, 10, and 20 (1 MΩ only) ±(0.057% of Input + 0.006% of FS + 900 µV) per °C
AC Amplitude Accuracy 50 Ω 1 MΩ Within ±5 °C of self-calibration temperature
±0.06 dB (±0.7%) at 50 kHz ±0.09 dB (±1.0%) at 50 kHz
Crosstalk, Typical ≤ –85 dB at 10 MHz CH 0 to/from CH 1, External Trigger to CH 0 or CH 1
Sparkle Code Rate, Typical

<300 ppt* with onboard clock or 200 MHz external clock

<3 ppt* with 150 MHz external clock

0 with 100 MHz external clock

Results based on 2 × 1012 samples

* ppt = parts per trillion (1012)

Bandwidth and Transient Response
Bandwidth (–3 dB) Range (Vpk-pk) 50 Ω 1 MΩ

Filters off

135 MHz above 40 °C

All ranges except 0.2 150 MHz 145 MHz up to 40 °C
0.2 85 MHz 75 MHz
Rise/Fall Time, Typical Range (Vpk-pk) 50 Ω and 1 MΩ Filters off
All ranges except 0.2 2.4 ns
0.2 3.3 ns
Bandwidth Limit Filters Noise Filter Antialias Filter Only one filter can be enabled at any given time. The antialias filter is enabled by default.
20 MHz, typical 2-pole Bessel filter 60 MHz, typical 4-pole elliptical filter
AC Coupling Cutoff (–3 dB) 12 Hz AC coupling available on 1 MΩ only
Passband Flatness Filter Settings Range (Vpk-pk) 50 Ω and 1 MΩ Referenced to 50 kHz
Filters Off All ranges except 0.2

±0.5 dB DC to 20 MHz

±1.0 dB 20 MHz to 50 MHz

±1.7 dB 50 MHz to 100 MHz

0.2

±0.6 dB DC to 20 MHz

±1.5 dB 20 MHz to 40 MHz

Antialias Filter On All ranges –1.0 dB to +2.0 dB DC to 55 MHz
NI 5124 Frequency Response (Typical)
Specification Value Comments
Spectral Characteristics
Spurious-Free Dynamic Range with Harmonics (SFDR), (Physical), Typical Range (Vpk-pk) 50 Ω 1 MΩ

Filters off or antialias filter on

10 MHz, –1 dBFS input signal

Includes the 2nd through the 5th harmonics

Measured from DC to 100 MHz on NI PXI-5124

Measured from 5 kHz to 100 MHz on NI PCI-5124

0.2 75 dBc 70 dBc
0.4 75 dBc 70 dBc
1 72 dBc 70 dBc
2 72 dBc 70 dBc
4  65 dBc  67 dBc
10  65 dBc  60 dBc
20
(1 MΩ only)
 60 dBc
Total Harmonic Distortion (THD), Typical Range (Vpk-pk) 50 Ω 1 MΩ

Filters off or antialias filter on

10 MHz, –1 dBFS input signal

Includes the 2nd through the 5th harmonics

0.2 –74 dBc – 68 dBc
0.4 –74 dBc – 68 dBc
1 –72 dBc – 68 dBc
2 –72 dBc – 67 dBc
4 – 63 dBc – 66 dBc
10 – 63 dBc –58 dBc
20
(1 MΩ only)
–58 dBc
Intermodulation Distortion, Typical 0.2 Vpk-pk to 2.0 Vpk-pk Ranges on 50 Ω Input

Filters off or antialias filter on

Two tones at 10.2 MHz and 11.2 MHz

Each tone is –7 dBFS

–75 dBc
Signal-to-Noise Ratio (SNR), Typical Range (Vpk-pk) 50 Ω 1 MΩ

Excludes harmonics

10 MHz, –1 dBFS input signal

Measured from DC to 100 MHz

Filters Off Antialias Filter On Filters Off Antialias Filter On
0.2 57 dB 56 dB 53 dB 55 dB
0.4 58 dB 57 dB 55 dB 57 dB
1 58 dB 58 dB 57 dB 57 dB
2 58 dB 58 dB 57 dB 57 dB
4 56 dB 58 dB
Signal to Noise and Distortion (SINAD), Typical Range (Vpk-pk) 50 Ω 1 MΩ

Includes harmonics

10 MHz, –1 dBFS input signal

Measured from DC to 100 MHz

Filters Off Antialias Filter On Filters Off Antialias Filter On
0.2 57 dB 56 dB 53 dB 55 dB
0.4 58 dB 57 dB 55 dB 57 dB
1 58 dB 58 dB 57 dB 57 dB
2 58 dB 58 dB 57 dB 57 dB
4 56 dB 57 dB
NI 5124 Dynamic Performance, 50 Ω , 1 Vpk-pk Range, 262,144 Point FFT (Typical)
Specification Value Comments
RMS Noise Range (Vpk-pk) 50 Ω 1 MΩ

Noise filter on

50 Ω terminator connected to input

0.2

NI PXI-5124: 94 μVrms (0.047% FS)

NI PCI-5124: 106 μVrms (0.053% FS)

NI PXI-5124: 104 μVrms (0.052% FS)

NI PCI-5124: 116 μVrms (0.058% FS)

0.4 188 μVrms (0.047% FS) 192 μVrms (0.048% FS)
1 470 μVrms (0.047% FS) 480 μVrms (0.048% FS)
2 940 μVrms (0.047% FS) 960 μVrms (0.048% FS)
4 1.88 mVrms (0.047% FS) 1.92 mVrms (0.048% FS)
10 4.7 mVrms (0.047% FS) 4.8 mVrms (0.048% FS)
20 (1 MΩ only) 9.4 mVrms (0.047% FS)
RMS Noise Range (Vpk-pk) 50 Ω 1 MΩ

Antialias filter on

50 Ω terminator connected to input

0.2

NI PXI-5124: 112 μVrms (0.056% FS)

NI PCI-5124: 126 μVrms (0.063% FS)

NI PXI-5124: 130 μVrms (0.065% FS)

NI PCI-5124: 146 μVrms (0.073% FS)

0.4 200 μVrms (0.05% FS) 216 μVrms (0.054% FS)
1 500 μVrms (0.05% FS) 510 μVrms (0.051% FS)
2 1.0 mVrms (0.05% FS) 1.02 mVrms (0.051% FS)
4 2.04 mVrms (0.051% FS) 2.16 mVrms (0.054% FS)
10 5.1 mVrms (0.051% FS) 5.2 mVrms (0.052% FS)
20 (1 MΩ only) N/A 10.2 mVrms (0.051% FS)
RMS Noise Range (Vpk-pk) 50 Ω 1 MΩ

Filters off

50 Ω terminator connected to input

0.2

NI PXI-5124: 114 μVrms (0.057% FS)

NI PCI-5124: 128 μVrms (0.064% FS)

NI PXI-5124: 164 μVrms (0.082% FS)

NI PCI-5124: 184 μVrms (0.092% FS)

0.4 204 μVrms (0.051% FS) 264 μVrms (0.066% FS)
1 510 μVrms (0.051% FS) 550 μVrms (0.055% FS)
2 1.02 mVrms (0.051% FS) 1.08 mVrms (0.054% FS)
4 2.08 mVrms (0.052% FS) 2.6 mVrms (0.065% FS)
10 5.2 mVrms (0.052% FS) 5.5 mVrms (0.055% FS)
20 10.6 mVrms (0.053% FS)
Representation of NI 5124 Spectral Noise Density on 0.2 V Range, Noise Filter Enabled, 1 MΩ Input Impedance
Representation of NI 5124 Spectral Noise Density on 0.2 V Range, Full Bandwidth, 50 Ω Input Impedance
Horizontal
Sample Clock
Specification Value Comments
Sources NI PXI-5124 NI PCI-5124 * Internal Sample Clock is locked to the Reference Clock or derived from the onboard VCXO

Internal, Onboard Clock (internal VCXO)*

External, CLK IN (front panel SMB connector)

External, PXI Star Trigger (backplane connector)

Internal, Onboard Clock (internal VCXO)*

External, CLK IN (front panel SMB connector)

Onboard Clock (Internal VCXO)
Sample Rate Range Real-Time Sampling (Single Shot) Random Interleaved Sampling (RIS)

Divide by n decimation used for all rates less than 200 MS/s

For more information about Sample Clock and decimation, refer to the NI High-Speed Digitizers Help.

3.052 kS/s to 200 MS/s 400 MS/s to 4 GS/s in multiples of 200 MS/s
Phase Noise Density, Typical

<–100 dBc/Hz at 100 Hz

<–120 dBc/Hz at 1 kHz

<–130 dBc/Hz at 10 kHz

10 MHz input signal
Sample Clock Jitter, Typical

1 ps rms (100 Hz to 100 kHz)

2 ps rms (100 Hz to 1 MHz)

Includes the effects of the converter aperture uncertainty and the clock circuitry jitter

Excludes trigger jitter

Timebase Frequency 200 MHz
Timebase Accuracy Not Phase-Locked to Reference Clock Phase-Locked to Reference Clock ppm = parts per million (1 × 10–6)
±25 ppm Equal to the Reference Clock accuracy
Sample Clock Delay Range ±1 Sample Clock period
Sample Clock Delay/Adjustment Resolution 5 ps
External Sample Clock
Sources NI PXI-5124 NI PCI-5124

CLK IN (front panel SMB connector)

PXI Star Trigger (backplane connector)

CLK IN (front panel SMB connector)
Frequency Range

50 MHz to 210 MHz (CLK IN)

50 MHz to 80 MHz (PXI Star Trigger)

Divide by n decimation available where 1 n 65,535

For more information about Sample Clock and decimation, refer to the NI High-Speed Digitizers Help.

Duty Cycle Tolerance 45% to 55%
Exported Reference Clock Destinations NI PXI-5124 NI PCI-5124

CLK OUT (front panel SMB connector)

PFI<0..1> (front panel 9-pin mini-circular DIN connector)

PXI_Trig<0..7> (backplane connector)

CLK OUT (front panel SMB connector)

PFI<0..1> (front panel 9-pin mini-circular DIN connector)

RTSI<0..7>

Sample Clock Exporting
Exported Sample Clock Destinations Destination Maximum Frequency * Decimated Sample Clock only
CLK OUT (front panel SMB connector) 210 MHz  
PXI_Trig<0..6> (backplane connector)* 20 MHz
PFI<0..1> (front panel 9-pin mini-circular DIN connector)* 25 MHz
RTSI<0..6>* 20 MHz
Phase-Locked Loop (PLL) Reference Clock
Specification Value Comments
Sources NI PXI-5124 NI PCI-5124

PXI_CLK10 (backplane connector)

CLK IN (front panel SMB connector)

RTSI  7

CLK IN (front panel SMB connector)

Frequency Range

1 MHz to 20 MHz in 1 MHz increments. Default of 10 MHz.

The PLL Reference Clock frequency must be accurate to ±50 ppm.

Duty Cycle Tolerance 45% to 55%
Exported Reference Clock Destinations NI PXI-5124 NI PCI-5124

CLK OUT (front panel SMB connector)

PFI<0..1> (front panel 9-pin mini-circular DIN connector)

PXI_Trig<0..6> (backplane connector)

CLK OUT (front panel SMB connector)

PFI<0..1> (front panel 9-pin mini-circular DIN connector)

RTSI<0..7>

CLK IN (Sample Clock and Reference Clock Input, Front Panel Connector)
Specification Value Comments
Input Voltage Range

Sine wave: 0.65 Vpk-pk to 2.8 Vpk-pk (0 dBm to 13 dBm)

Square wave: 0.2 Vpk-pk to 2.8 Vpk-pk

Maximum Input Overload 7 Vrms with |Peaks| 10 V
Impedance 50 Ω
Coupling AC
CLK OUT (Sample Clock and Reference Clock Output, Front Panel Connector)
Specification Value Comments
Output Impedance 50 Ω
Logic Type 3.3 V CMOS
Maximum Drive Current ±48 mA
Trigger
Reference (Stop) Trigger
Specification Value Comments
Trigger Types and Sources Types Sources Refer to the following sections and the NI High-Speed Digitizers Help for more information about what sources are available for each trigger type.
Edge, Window, Hysteresis, Video, Digital, Immediate, and Software CH 0, CH 1, TRIG, PXI_Trig<0..6>, PFI<0..1>, PXI Star Trigger, Software, and RTSI<0..6>
Time Resolution TDC Onboard Clock External Clock TDC = Time to Digital Conversion Circuit
On 50 ps N/A
Off 5 ns External Clock Period
Minimum Rearm Time TDC Rearm Time Holdoff set to 0. Onboard sample clock at maximum rate.
On 10 μs
Off μs
Holdoff TDC Onboard Clock External Clock
On 10 μs to 85.899 s N/A
Off μs to 85.899 s 200 × (External Clock Period) to (232 – 1) × (External Clock Period)
Analog Trigger (Edge, Window, and Hysteresis Trigger Types)
Sources

CH 0 (front panel BNC connector)

CH 1 (front panel BNC connector)

TRIG (front panel BNC connector)

Trigger Level Range CH 0, CH 1 TRIG (External Trigger)
100% FS ±5 V
Trigger Level Resolution 10 bits (1 in 1,024)
Edge Trigger Sensitivity CH 0, CH 1 TRIG (External Trigger)
3.5% FS up to 50 MHz, increasing to 10% FS at 150 MHz 0.25 Vpk-pk up to 100 MHz, increasing to 1 Vpk-pk at 200 MHz
Level Accuracy, Typical CH 0, CH 1 TRIG (External Trigger)
±4.7% FS up to 10 MHz ±0.35 V up to 10 MHz
Trigger Jitter 80 ps rms Within ±5 °C of self-calibration temperature
Trigger Filters Low-Frequency (LF) Reject High-Frequency (HF) Reject
50 kHz 50 kHz
Digital Trigger (Digital Trigger Type)
Sources NI PXI-5124 NI PCI-5124

PXI_Trig<0..6> (backplane connector)

PFI<0..1> (front panel SMB connector)

PXI Star Trigger (backplane connector)

RTSI<0..6>

PFI<0..1> (front panel SMB connector)

Video Trigger (Video Trigger Type)
Sources

CH 0 (front panel BNC connector)

CH 1 (front panel BNC connector)

TRIG (front panel BNC connector)

Types

Specific Line

Any Line

Specific Field

Standard Negative sync of NTSC, PAL, or SECAM signal
TRIG (External Trigger, Front Panel Connector)
Specification Value Comments
Connector BNC
Impedance 1 MΩ in parallel with 22 pF
Coupling AC, DC
AC-Coupling Cutoff (–3 dB) 12 Hz
Input Voltage Range ±5 V
Maximum Input Overload |Peaks| 42 V
PFI 0 and PFI 1 (Programmable Function Interface, AUX Front Panel Connector)
Specification Value Comments
Connector 9-pin mini-circular DIN
Direction Bi-directional
As an Input (Trigger)
Destinations

Start Trigger (Acquisition Arm)

Reference (Stop) Trigger

Arm Reference Trigger

Advance Trigger

Input Impedance 150 kΩ
VIH 2.0 V
VIL 0.8 V
Maximum Input Overload – 0.5 V, 5.5 V
Maximum Frequency 25 MHz
As an Output (Event)
Sources

Start Trigger (Acquisition Arm)

Reference (Stop) Trigger

End of Record

Done (End of Acquisition)

Probe Compensation (1 kHz, 50% duty cycle square wave, PFI 1 only)

Output Impedance 50 Ω
Logic Type 3.3 V CMOS
Maximum Drive Current ±24 mA
Maximum Frequency 25 MHz
TClk Specifications

National Instruments TClk synchronization method and the NI-TClk driver are used to align the sample clocks on any number of SMC-based modules in a chassis. For more information about TClk synchronization, refer to the NI-TClk Synchronization Help, which is located within the NI High-Speed Digitizers Help.

  • Specifications are valid for any number of modules installed in one NI PXI-1042 chassis.
  • All parameters set to identical values for each SMC-based module.
  • Sample Clock set to200 MS/s and all filters are disabled.
  • For other configurations, including multichassis systems, contact NI Technical Support at ni.com/support.

Note Although you can use NI-TClk to synchronize nonidentical modules, these specifications apply only to synchronizing identical modules.

Specification Value Comments
Intermodule SMC Synchronization Using NI-TClk for Identical Modules (Typical)
Skew 500 ps

Caused by clock and analog path delay differences

No manual adjustment performed

Average Skew After Manual Adjustment <10 ps For information about manual adjustment, refer to the Synchronization Repeatability Optimization topic in the NI-TClk Synchronization Help. For additional help with the adjustment process, contact NI Technical Support at ni.com/support.
Sample Clock Delay/Adjustment Resolution 5 ps
Waveform Specifications
Specification Value Comments
Onboard Memory Size 8 MB per channel standard (4 megasamples per channel) * NI PXI-5124 only
32 MB per channel option (16 megasamples per channel)
256 MB per channel option (128 megasamples per channel)
512 MB per channel option (256 megasamples per channel)*
Minimum Record Length 1 Sample
Number of Pretrigger Samples Zero up to full Record Length Single-record mode and multiple-record mode
Number of Posttrigger Samples Zero up to full Record Length Single-record mode and multiple-record mode
Maximum Number of Records in Onboard Memory 8 MB/channel 21,845

* NI PXI-5124 only

It is possible to exceed these numbers if you fetch records while acquiring data.

For more information, refer to the NI High-Speed Digitizers Help.

32 MB/channel 87,381
256 MB/channel 100,000
512 MB/channel* 100,000
Allocated Onboard Memory per Record

(Record Length × 2 bytes/S) + 200 bytes, rounded up to next multiple of 128 bytes

or

384 bytes, whichever is greater

Calibration
Specification Value Comments
Self-Calibration Self-calibration is done on software command. The calibration corrects for gain, offset, frequency response, triggering, and timing adjustment errors for all input ranges.
External Calibration (Factory Calibration) The external calibration calibrates the VCXO and the voltage reference. Appropriate constants are stored in nonvolatile memory.
Interval for External Calibration 2 years
Warm-Up Time 15 minutes
Power
Specification Typical Value Comments
  NI PXI-5124 NI PCI-5124
+3.3 VDC 1.3 A 1.3 A
+5 VDC 1.7 A 2.7 A
+12 VDC 130 mA 130 mA
–12 VDC 270 mA 0 A
Total Power 17.6 W 19.4 W
Software
Specification Value Comments
Driver Software

NI-SCOPE 2.7 or later

NI-SCOPE is an IVI-compliant driver that allows you to configure, control, and calibrate the NI 5124. NI-SCOPE provides application programming interfaces for many development environments.

Application Software

NI-SCOPE provides programming interfaces, documentation, and examples for the following application development environments:

  • LabVIEW
  • LabWindows/CVI
  • Measurement Studio
  • Microsoft Visual C/C++
  • Microsoft Visual Basic
Interactive Soft Front Panel and Configuration

The Scope Soft Front Panel supports interactive control of the NI 5124. The Scope Soft Front Panel is included on the NI-SCOPE CD.

National Instruments Measurement & Automation Explorer (MAX) also provides interactive configuration and test tools for the NI 5124. MAX is included on the NI-SCOPE CD.

Environment
NI PXI-5124

Note To ensure that the NI PXI-5124 cools effectively, follow the guidelines in the Maintain Forced Air Cooling Note to Users included in the NI PXI-5124 kit. The NI PXI-5124 is intended for indoor use only.

Specification Value Comments
Operating Temperature 0 ºC to +55 ºC in all NI PXI chassis except the following: 0 ºC to +45 ºC when installed in an NI PXI-1000/B or PXI-101x chassis Meets IEC-60068-2-1 and IEC-60068-2-2
Storage Temperature – 40 ºC to +71 ºC Meets IEC-60068-2-1 and IEC-60068-2-2
Operating Relative Humidity

10% to 90%, noncondensing

Meets IEC-60068-2-56

Storage Relative Humidity

5% to 95%, noncondensing

Meets IEC-60068-2-56

Operating Shock

30 g, half-sine, 11 ms pulse

Meets IEC-60068-2-27

Test profile developed in accordance with MIL-PRF-28800F

Storage Shock

50 g, half-sine, 11 ms pulse

Meets IEC-60068-2-27

Test profile developed in accordance with MIL-PRF-28800F

Operating Vibration

5 Hz to 500 Hz, 0.31 grms

Meets IEC-60068-2-64

Storage Vibration

5 Hz to 500 Hz, 2.46 grms

Meets IEC-60068-2-64

Test profile exceeds requirements of MIL-PRF-28800F, Class 3

Altitude 2,000 m maximum (at 25 °C ambient temperature)
Pollution Degree 2
NI PCI-5124

Note To ensure that the NI PCI-5124 cools effectively, make sure that the chassis in which it is used has active cooling that provides at least some airflow across the PCI card cage. To maximize airflow and extend the life of the device, leave any adjacent PCI slots empty. Refer to the Maintain Forced Air Cooling Note to Users included in the NI PCI-5124 kit for important cooling information. The NI PCI-5124 is intended for indoor use only.

Specification Value Comments
Operating Temperature

0 ºC to +45 ºC

Meets IEC-60068-2-1 and IEC-60068-2-2

Storage Temperature

– 40 ºC to +71 ºC

Meets IEC-60068-2-1 and IEC-60068-2-2

Operating Relative Humidity

10% to 90%, noncondensing

Meets IEC-60068-2-56

Storage Relative Humidity

5% to 95%, noncondensing

Meets IEC-60068-2-56

Storage Shock

50 g, half-sine, 11 ms pulse

Meets IEC-60068-2-27

Test profile developed in accordance with MIL-PRF-28800F

Storage Vibration

5 Hz to 500 Hz, 2.46 grms

Meets IEC-60068-2-64

Test profile exceeds requirements of MIL-PRF-28800F, Class 3

Altitude 2,000 m maximum (at 25 °C ambient temperature)
Pollution Degree 2
Safety, Electromagnetic Compatibility, and CE Compliance
Safety Standards
This product is designed to meet the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use:
  • IEC 61010-1, EN 61010-1
  • UL 61010-1, CSA 61010-1

Note For UL and other safety certifications, refer to the product label or the Online Product Certification section.

Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
  • EN 61326 (IEC 61326): Class A emissions; Basic immunity
  • EN 55011 (CISPR 11): Group 1, Class A emissions
  • AS/NZS CISPR 11: Group 1, Class A emissions
  • FCC 47 CFR Part 15B: Class A emissions
  • ICES-001: Class A emissions

Note For the standards applied to assess the EMC of this product, refer to the Online Product Certification section.

Note For EMC compliance, operate this device with RG223/U or equivalent shielded cable. Operate according to product documentation.

CE Compliance
 This product meets the essential requirements of applicable European Directives, as amended for CE marking, as follows:
  • 2006/95/EC; Low-Voltage Directive (safety)
  • 2004/108/EC; Electromagnetic Compatibility Directive (EMC)
Online Product Certification
 Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by module number or product line, and click the appropriate link in the Certification column.
Environmental Management

National Instruments is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial not only to the environment but also to NI customers.

For additional environmental information, refer to the NI and the Environment Web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.

Waste Electrical and Electronic Equipment (WEEE)

EU Customers At the end of the product life cycle, all products must be sent to a WEEE recycling center. For more information about WEEE recycling centers, National Instruments WEEE initiatives, and compliance with WEEE Directive 2002/96/EC on Waste Electrical and Electronic Equipment, visit ni.com/environment/weee.htm.

 

 

Physical
Front Panel Connectors
Label Function Connector Type
CH 0 Analog Input BNC female
CH 1 Analog Input BNC female
TRIG External Trigger BNC female
CLK IN Sample Clock Input and Reference Clock Input SMB jack
CLK OUT Sample Clock Output and Reference Clock Output SMB jack
AUX I/O PFI 0, PFI 1 9-pin mini-circular DIN
NI PXI-5124 Front Panel Indicators
Label Function For more information, refer to the NI High-Speed Digitizers Help.
ACCESS The ACCESS LED indicates the status of the PCI bus and the interface from the NI PXI-5124 to the controller.
ACTIVE The ACTIVE LED indicates the status of the onboard acquisition hardware of the NI PXI-5124.
Dimensions and Weight
NI PXI-5124
Dimensions

3U, One slot, PXI/cPCI Module
21.6 × 2.0 × 13.0 cm
(8.5 × 0.8 × 5.1 in.)



Weight 383 g (13.5 oz)
NI PCI-5124
Dimensions

35.5 × 2.0 × 11.3 cm
(14.0 × 0.8 × 4.4 in.)



Weight 455 g (16 oz)

Back to Top