Last Revised: 2012-12-11 16:15:33.0
|
|
|
|
NI system assurance programs are designed to make it even easier for you to own an NI system. These programs include configuration and deployment services for your NI PXI, CompactRIO, or Compact FieldPoint system. The NI Basic System Assurance Program provides a simple integration test and ensures that your system is delivered completely assembled in one box. When you configure your system with the NI Standard System Assurance Program, you can select from available NI system driver sets and application development environments to create customized, reorderable software configurations. Your system arrives fully assembled and tested in one box with your software preinstalled. When you order your system with the standard program, you also receive system-specific documentation including a bill of materials, an integration test report, a recommended maintenance plan, and frequently asked question documents. Finally, the standard program reduces the total cost of owning an NI system by providing three years of warranty coverage and calibration service. Use the online product advisors at ni.com/advisor to find a system assurance program to meet your needs.
NI measurement hardware is calibrated to ensure measurement accuracy and verify that the device meets its published specifications. To ensure the ongoing accuracy of your measurement hardware, NI offers basic or detailed recalibration service that provides ongoing ISO 9001 audit compliance and confidence in your measurements. To learn more about NI calibration services or to locate a qualified service center near you, contact your local sales office or visit ni.com/calibration.
Get answers to your technical questions using the following National Instruments resources.
While you may never need your hardware repaired, NI understands that unexpected events may lead to necessary repairs. NI offers repair services performed by highly trained technicians who quickly return your device with the guarantee that it will perform to factory specifications. For more information, visit ni.com/repair.
The NI training and certification program delivers the fastest, most certain route to increased proficiency and productivity using NI software and hardware. Training builds the skills to more efficiently develop robust, maintainable applications, while certification validates your knowledge and ability.
NI offers options for extending the standard product warranty to meet the life-cycle requirements of your project. In addition, because NI understands that your requirements may change, the extended warranty is flexible in length and easily renewed. For more information, visit ni.com/warranty.
NI offers design-in consulting and product integration assistance if you need NI products for OEM applications. For information about special pricing and services for OEM customers, visit ni.com/oem.
Our Professional Services Team is comprised of NI applications engineers, NI Consulting Services, and a worldwide National Instruments Alliance Partner program of more than 700 independent consultants and integrators. Services range from start-up assistance to turnkey system integration. Visit ni.com/alliance.
|
8-Bit 250 MS/s Digitizer This document lists the specifications for the NI PXI/PCI-5114 (NI 5114) high-speed digitizer. Unless otherwise noted, these specifications are valid for the following conditions:
Typical values are representative of an average unit operating at room temperature. Specifications are subject to change without notice. For the most recent NI 5114 specifications, visit ni.com/manuals. To access the NI 5114 documentation, including the NI High-Speed Digitizers Getting Started Guide, which contains functional descriptions of the NI 5114 signals, navigate to Start»All Programs»National Instruments»NI-SCOPE»Documentation. Note If the NI 5114 has been in use, it may exceed safe handling temperatures and cause burns. Allow the NI 5114 to cool before removing it from the PXI chassis or PC. Refer to the Environment section for operating temperatures of this device. |
||
| Vertical | ||
| Analog Input (Channel 0 and Channel 1) |
| Specification | Value | Comments | |||||
|---|---|---|---|---|---|---|---|
| Number of Channels | Two (simultaneously sampled) | — | |||||
| Connector | BNC | — | |||||
| Impedance and Coupling | |||||||
| Input Impedance |
50 Ω ±1.5% 1 MΩ ±1% in parallel with a typical capacitance of 26 pF |
Software selectable | |||||
| Input Coupling | AC, DC, GND | AC coupling available on 1 MΩ only | |||||
| Voltage Levels | |||||||
| Full Scale (FS) Input Range and Programmable Vertical Offset | 50 Ω | 1 MΩ | — | ||||
| Range (Vpk-pk) | Vertical Offset Range (V) |
Range (Vpk-pk) | Vertical Offset Range (V) |
||||
| 0.04 | ±0.8 | 0.04 | ±0.8 | ||||
| 0.1 | ±0.8 | 0.1 | ±0.8 | ||||
| 0.2 | ±0.8 | 0.2 | ±0.8 | ||||
| 0.4 | ±0.8 | 0.4 | ±0.8 | ||||
| 1 | ±6.5 | 1.0 | ±8.0 | ||||
| 2 | ±6.0 | 2.0 | ±8.0 | ||||
| 4 | ±5.0 | 4.0 | ±8.0 | ||||
| 10 | ±2.0 | 10 | ±30 | ||||
| 20 | ±25 | ||||||
| 40 | ±15 | ||||||
| Maximum Input Overload | 50 Ω | 1 MΩ | — | ||||
| 7 Vrms with |Peaks| ≤10 V | |Peaks| ≤35 V | ||||||
| Accuracy | |||||||
| Resolution | 8 bits | — | |||||
| DC Accuracy (Programmable Vertical Offset = 0 V) |
NI PXI-5114: ±(1.5% of Input + 0.3% of FS + 200 μV) NI PCI-5114: ±(1.5% of Input + 0.3% of FS + 280 μV) |
Within ±5 °C of self-calibration temperature | |||||
| Programmable Vertical Offset Accuracy | ±2% of offset setting | Within ±5 °C of self-calibration temperature | |||||
| DC Drift | ±(0.03% of Input + 0.06% of FS + 40 μV) per °C | — | |||||
| Crosstalk, Typical |
≤–60 dB at 10 MHz ≤– 45 dB at 100 MHz |
CH 0 to/from CH 1, External Trigger to CH 0 or CH 1 | |||||
| Bandwidth and Transient Response | |||||||
| Bandwidth (–3 dB) | Range (Vpk-pk) | Bandwidth | Rise/Fall Time, Typical | — | |||
| All ranges except 0.04 | 125 MHz | 2.8 ns | |||||
| 0.04 | 100 MHz | 3.5 ns | |||||
| Bandwidth Limit Filter | 20 MHz Noise Filter | — | |||||
| AC Coupling* Cutoff (–3 dB), Typical | 12 Hz | * AC coupling available on 1 MΩ only | |||||
| Passband Flatness | ±1 dB up to 50 MHz |
Referenced to 50 kHz Bandwidth limit filter off |
|||||
| NI 5114 Frequency Response (Typical) | ||
![]() |
| Specification | Value | Comments | |||||
|---|---|---|---|---|---|---|---|
| Spectral Characteristics | |||||||
| Spurious-Free Dynamic Range with Harmonics (SFDR), Typical | Range (Vpk-pk) |
10 MHz, –1 dBFS input signal Includes the 2nd through the 5th harmonics Measured from DC to 125 MHz 20 MHz bandwidth limit filter off |
|||||
| All ranges except 0.04 | 0.04 | ||||||
| 58 dBc | 58 dBc | ||||||
| Total Harmonic Distortion (THD), Typical | –58 dBc | –58 dBc | |||||
| Effective Number of Bits (ENOB), Calculated* | 7.2 | 6.2 | |||||
| Signal to Noise and Distortion (SINAD), Typical | 44 dB | 38 dB | |||||
| RMS Noise | Range (Vpk-pk) | 20 MHz Filter On | 20 MHz Filter Off | 50 Ω terminator connected to input | |||
| All ranges except 0.04 | 0.28% FS | 0.28% FS | |||||
| 0.04 | 0.28% FS | 0.45% FS | |||||
|
*ENOB = log2(sinad) – ½log2(1.5) – log2(A/V) where sinad = the linear representation of SINAD Refer to 1057-1994 IEEE Standard for Digitizing Waveform Recorders for information on equation derivation. |
||
| Horizontal | ||
| Sample Clock |
| Specification | Value | Comments | |||||
|---|---|---|---|---|---|---|---|
| Sources |
Internal, Onboard Clock (internal VCXO)* External, CLK IN (front panel SMB connector) |
*Internal Sample Clock is locked to the Reference Clock or derived from the onboard VCXO | |||||
| Onboard Clock (Internal VCXO) | |||||||
| Sample Rate Range | Real-Time Sampling (Single Shot) | Random Interleaved Sampling (RIS) |
† Divide by n decimation used for all rates less than 250 MS/s For more information about Sample Clock and decimation, refer to the NI High-Speed Digitizers Help. |
||||
| 3.815 kS/s to 250MS/s† | 250 MS/s to 5 GS/s in increments of 250 MS/s | ||||||
| Timebase Frequency | 250 MHz | When not using External Sample Clock | |||||
| Timebase Accuracy | Not Phase-Locked to Reference Clock | Phase-Locked to Reference Clock | ppm = parts per million (1 × 10–6) | ||||
| ±25 ppm | Equal to the Reference Clock accuracy | ||||||
| Sample Clock Delay Range | ±1 Sample Clock period | — | |||||
| Sample Clock Delay Resolution | ≤20 ps | — | |||||
| External Sample Clock | |||||||
| Sources | CLK IN (front panel SMB connector) | — | |||||
| Frequency Range | 50 MHz to 250 MHz |
Divide by n decimation available where 1 ≤ n ≤ 65,535 For more information about Sample Clock and decimation, refer to the NI High-Speed Digitizers Help. |
|||||
| Duty Cycle Tolerance | 45% to 55% | — | |||||
| Phase-Locked Loop (PLL) Reference Clock |
| Specification | Value | |||||
|---|---|---|---|---|---|---|
| Sources | NI PXI-5114 | NI PCI-5114 | ||||
|
PXI_CLK10 (backplane connector) CLK IN (front panel SMB connector) |
RTSI 7 CLK IN (front panel SMB connector) |
|||||
| Frequency Range |
1 MHz to 20 MHz in 1 MHz increments Default of 10 MHz The PLL Reference Clock frequency must be accurate to ±50 ppm |
|||||
| Duty Cycle Tolerance | 45% to 55% | |||||
| Exported Reference Clock Destinations | NI PXI-5114 | NI PCI-5114 | ||||
|
PFI <0..1> (front panel 9-pin mini-circular DIN connector) PXI_Trig <0..7> (backplane connector) |
PFI <0..1> (front panel 9-pin mini-circular DIN connector) RTSI <0..7> |
|||||
| CLK IN (Sample Clock and Reference Clock Input, Front Panel Connector) |
| Specification | Value | |||||
|---|---|---|---|---|---|---|
| Input Voltage Range | Sine wave: 0.65 Vpk-pk to 2.8 Vpk-pk (0 dBm to 13 dBm) Square wave: 0.2 Vpk-pk to 2.8 Vpk-pk | |||||
| Maximum Input Overload | 7 Vrms with |Peaks| ≤10 V | |||||
| Impedance | 50 Ω | |||||
| Coupling | AC | |||||
| Trigger | ||
| Reference (Stop) Trigger |
| Specification | Value | Comments | |||||
|---|---|---|---|---|---|---|---|
| Trigger Types and Sources | Types | Sources | Refer to the following sections and to NI High-Speed Digitizers Help for more information. | ||||
| Edge, Window, Hysteresis, Video, Digital, Immediate, and Software | CH 0, CH 1, TRIG, PXI_Trig<0..6>, PFI <0..1>, PXI Star Trigger, RTSI<0..6>, and Software | ||||||
| Time Resolution | TDC | Onboard Clock | External Clock | TDC = Time to Digital Conversion Circuit | |||
| On | 40 ps | N/A | |||||
| Off | 4 ns | External Clock Period | |||||
| Minimum Rearm Time | TDC | Rearm Time | Holdoff set to 0. Onboard sample clock at maximum rate. | ||||
| On | 10 μs | ||||||
| Off | 2 μs | ||||||
| Holdoff | From Rearm Time up to [(235 – 1) × (Sample Clock Period)] | — | |||||
| Trigger Delay | From 0 up to [(235 – 1) – posttrigger samples] × (1/sample rate), in seconds | — | |||||
| Analog Trigger (Edge, Window, and Hysteresis Trigger Types) | |||||||
| Sources |
CH 0 (front panel BNC connector) CH 1 (front panel BNC connector) TRIG (front panel BNC connector) |
— | |||||
| Trigger Level Resolution | 8 bits (1 in 256) | ||||||
| Trigger Level Range | CH 0, CH 1 | TRIG (External Trigger) | — | ||||
| 100% FS | ±5 V | ||||||
| Edge Trigger Sensitivity | 5% FS up to 100 MHz | 0.5 Vpk-pk up to 100 MHz | |||||
| Level Accuracy, Typical | ±5% FS up to 10 MHz | ±0.5 V up to 10 MHz | |||||
| Jitter | ≤65 ps rms | — | |||||
| Trigger Filters | Low Frequency (LF) Reject | High Frequency (HF) Reject | — | ||||
| 50 kHz | 50 kHz | ||||||
| Digital Trigger (Digital Trigger Type) | |||||||
| Sources | NI PXI-5114 | NI PCI-5114 | — | ||||
|
PXI_Trig <0..6> PFI <0..1> PXI Star Trigger |
RTSI <0..6> PFI <0..1> |
||||||
| Video Trigger (Video Trigger Type) | |||||||
| Sources |
CH 0 (front panel BNC connector) CH 1 (front panel BNC connector) TRIG (front panel BNC connector) |
— | |||||
| Types |
Specific Line Any Line Specific Field |
— | |||||
| Standards |
SDTV: M-NTSC, B/G-PAL, SECAM, M-PAL EDTV: 480i/59.94 fps, 480i/60 fps, 480p/59.94 Fps, 480p/60 Fps, 576i/50 fps, 576p/50 Fps HDTV: 720p/50 Fps, 720p/59.94 Fps, 720p/60 Fps, 1080i/50 fps, 1080i/59.94 fps, 1080i/60 fps, 1080p/24 Fps |
fps = fields per second Fps = Frames per second |
|||||
| TRIG (External Trigger, Front Panel Connector) |
| Specification | Value | |||||
|---|---|---|---|---|---|---|
| Connector | BNC | |||||
| Impedance | 1 MΩ in parallel with 22 pF | |||||
| Coupling | AC, DC | |||||
| AC-Coupling Cutoff (–3 dB) | 12 Hz | |||||
| Input Voltage Range | ±5 V | |||||
| Maximum Input Overload | |Peaks| ≤42 V | |||||
| PFI 0 and PFI 1 (Programmable Function Interface, AUX Front Panel Connectors) |
| Specification | Value | |||||
|---|---|---|---|---|---|---|
| Connector | 9-pin mini-circular DIN | |||||
| Direction | Bi-directional | |||||
| As an Input (Trigger) | ||||||
| Destinations |
Start Trigger (Acquisition Arm) Reference (Stop) Trigger Arm Reference Trigger Advance Trigger |
|||||
| Input Impedance | 150 kΩ | |||||
| VIH | 2.0 V | |||||
| VIL | 0.8 V | |||||
| Maximum Input Overload | –0.5 V, 5.5 V | |||||
| Maximum Frequency | 25 MHz | |||||
| As an Output (Event) | ||||||
| Sources |
Start Trigger (Acquisition Arm) Reference (Stop) Trigger End of Record Done (End of Acquisition) Probe Compensation (1 kHz, 50% duty cycle square wave, PFI 1 only) |
|||||
| Output Impedance | 50 Ω | |||||
| Logic Type | 3.3 V CMOS | |||||
| Maximum Drive Current | ±24 mA | |||||
| Maximum Frequency | 25 MHz | |||||
| TClk Specifications | ||
|
National Instruments TClk synchronization method and the NI-TClk driver are used to align the sample clocks on any number of SMC-based modules in a chassis. For more information about TClk synchronization, refer to the NI-TClk Synchronization Help, which is located within the NI High-Speed Digitizers Help.
Note Although you can use NI-TClk to synchronize nonidentical modules, these specifications apply only to synchronizing identical modules. |
| Specification | Value | Comments | |||||
|---|---|---|---|---|---|---|---|
| Intermodule SMC Synchronization Using NI-TClk for Identical Modules (Typical) | |||||||
| Skew | 500 ps |
Caused by clock and analog path delay differences No manual adjustment performed |
|||||
| Average Skew After Manual Adjustment | <20 ps | For information about manual adjustment, refer to the Synchronization Repeatability Optimization topic in the NI-TClk Synchronization Help. For additional help with the adjustment process, contact NI Technical Support at ni.com/support. | |||||
| Sample Clock Adjustment Resolution | <20 ps | — | |||||
| Waveform Specifications |
| Specification | Value | Comments | |||||
|---|---|---|---|---|---|---|---|
| Onboard Memory Size | 8 MB per Channel Standard | 8 megasamples per channel | — | ||||
| 64 MB per Channel Option | 64 megasamples per channel | ||||||
| 256 MB per Channel Option | 256 megasamples per channel | ||||||
| Minimum Record Length | 1 Sample | — | |||||
| Number of Pretrigger Samples | Zero up to full Record Length | Single-record mode and multiple-record mode | |||||
| Number of Posttrigger Samples | Zero up to full Record Length | Single-record mode and multiple-record mode | |||||
| Maximum Number of Records in Onboard Memory | 8 MB/channel | 32,768 | * It is possible to exceed these numbers if you fetch records while acquiring data. For more information, refer to the NI High-Speed Digitizers Help. | ||||
| 64 MB/channel | 100,000* | ||||||
| 256 MB/channel | 100,000* | ||||||
| Allocated Onboard Memory per Record |
(Record Length × 1 byte/S) + 240 bytes, rounded up to next multiple of 128 bytes or 256 bytes, whichever is greater |
— | |||||
| Calibration |
| Specification | Value | |||||
|---|---|---|---|---|---|---|
| Self-Calibration | Self-calibration is done on software command. The calibration corrects for gain, offset, compensated 1 MΩ attenuator, triggering, and timing adjustment errors for all input ranges. | |||||
| External Calibration (Factory Calibration) | The external calibration calibrates the VCXO, gain, and the voltage reference. Appropriate constants are stored in nonvolatile memory. | |||||
| Interval for External Calibration | 2 years | |||||
| Warm-Up Time | 15 minutes | |||||
| Power |
| Specification | Typical Value | |||||
|---|---|---|---|---|---|---|
| +3.3 VDC | NI PXI-5114 | NI PCI-5114 | ||||
| 840 mA | 1.6 A | |||||
| +5 VDC | 1.1 A | 1.7 A | ||||
| +12 VDC | 250 mA | 45 mA | ||||
| –12 VDC | 170 mA | — | ||||
| Total Power | 13.32 W | 14.32 W | ||||
| Software |
| Specification | Value | |||||
|---|---|---|---|---|---|---|
| Driver Software |
NI PXI-5114: NI-SCOPE 2.9 or later NI PCI-5114: NI-SCOPE 3.1 or later NI-SCOPE is an IVI-compliant driver that allows you to configure, control, and calibrate the NI 5114. NI-SCOPE provides application programming interfaces for many development environments. |
|||||
| Application Software |
NI-SCOPE provides programming interfaces, documentation, and examples for the following application development environments:
|
|||||
| Interactive Soft Front Panel and Configuration |
The Scope Soft Front Panel 2.3 or later supports interactive control of the NI 5114. The Scope Soft Front Panel is included on the NI-SCOPE CD. National Instruments Measurement & Automation Explorer (MAX) also provides interactive configuration and test tools for the NI 5114. MAX is included on the NI-SCOPE CD. |
|||||
| Environment | ||
| NI PXI-5114 | ||
|
Note To ensure that the NI PXI-5114 cools effectively, follow the guidelines in the Maintain Forced-Air Cooling Note to Users included in the NI PXI-5114 kit. The NI PXI-5114 is intended for indoor use only. |
| Specification | Value | |||||
|---|---|---|---|---|---|---|
| Operating Temperature |
0 ºC to +55 ºC in all NI PXI chassis except the following: 0 ºC to +45 ºC when installed in an NI PXI-1000/B or PXI-101x chassis Meets IEC-60068-2-1 and IEC-60068-2-2 |
|||||
| Storage Temperature |
– 40 ºC to +71 ºC Meets IEC-60068-2-1 and IEC-60068-2-2 |
|||||
| Operating Relative Humidity |
10% to 90%, noncondensing Meets IEC-60068-2-56 |
|||||
| Storage Relative Humidity |
5% to 95%, noncondensing Meets IEC-60068-2-56 |
|||||
| Operating Shock |
30 g, half-sine, 11 ms pulse Meets IEC-60068-2-27 Test profile developed in accordance with MIL-PRF-28800F |
|||||
| Storage Shock |
50 g, half-sine, 11 ms pulse Meets IEC-60068-2-27 Test profile developed in accordance with MIL-PRF-28800F |
|||||
| Operating Vibration |
5 Hz to 500 Hz, 0.31 grms Meets IEC-60068-2-64 |
|||||
| Storage Vibration |
5 Hz to 500 Hz, 2.46 grms Meets IEC-60068-2-64 Test profile exceeds requirements of MIL-PRF-28800F, Class 3 |
|||||
| Altitude | 2,000 m maximum (at 25 °C ambient temperature) | |||||
| Pollution Degree | 2 | |||||
| NI PCI-5114 | ||
|
Note To ensure that the NI PCI-5114 cools effectively, make sure that the chassis in which it is used has active cooling that provides at least some airflow across the PCI card cage. To maximize airflow and extend the life of the device, leave any adjacent PCI slots empty. Refer to the Maintain Forced-Air Cooling Note to Users included in the NI PCI-5114 kit for important cooling information. The NI PCI-5114 is intended for indoor use only. |
| Specification | Value | |||||
|---|---|---|---|---|---|---|
| Operating Temperature |
0 ºC to +45 ºC Meets IEC-60068-2-1 and IEC-60068-2-2 |
|||||
| Storage Temperature |
– 40 ºC to +71 ºC Meets IEC-60068-2-1 and IEC-60068-2-2 |
|||||
| Operating Relative Humidity |
10% to 90%, noncondensing Meets IEC-60068-2-56 |
|||||
| Storage Relative Humidity |
5% to 95%, noncondensing Meets IEC-60068-2-56 |
|||||
| Storage Shock |
50 g, half-sine, 11 ms pulse Meets IEC-60068-2-27 Test profile developed in accordance with MIL-PRF-28800F |
|||||
| Storage Vibration |
5 Hz to 500 Hz, 2.46 grms Meets IEC-60068-2-64 Test profile exceeds requirements of MIL-PRF-28800F, Class 3 |
|||||
| Altitude | 2,000 m maximum (at 25 °C ambient temperature) | |||||
| Pollution Degree | 2 | |||||
| Safety, Electromagnetic Compatibility, and CE Compliance | ||
| Safety Standards | ||
|
This product is designed to meet the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use:
Note For UL and other safety certifications, refer to the product label or the Online Product Certification section. |
||
| Electromagnetic Compatibility | ||
|
This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use:
Note For the standards applied to assess the EMC of this product, refer to the Online Product Certification section. Note For EMC compliance, operate this device with RG223/U or equivalent shielded cable. Operate according to product documentation. |
||
| CE Compliance | ||
| Online Product Certification | ||
| Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by module number or product line, and click the appropriate link in the Certification column. | ||
| Environmental Management | ||
|
National Instruments is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial not only to the environment but also to NI customers. For additional environmental information, refer to the NI and the Environment Web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document. |
||
| Waste Electrical and Electronic Equipment (WEEE) | ||
|
EU Customers At the end of the product life cycle, all products must be sent to a WEEE recycling center. For more information about WEEE recycling centers, National Instruments WEEE initiatives, and compliance with WEEE Directive 2002/96/EC on Waste Electrical and Electronic Equipment, visit ni.com/environment/weee.htm. |
||
|
|
||
| Physical | ||
| Front Panel Connectors |
| Label | Function | Connector Type | Comments | ||||
| CH 0 | Analog Input | BNC female | — | ||||
| CH 1 | Analog Input | BNC female | |||||
| TRIG | External Trigger | BNC female | |||||
| CLK IN | Sample Clock Input and Reference Clock Input | SMB jack | |||||
| AUX I/O | PFI 0, PFI 1 | 9-pin mini-circular DIN | |||||
| NI PXI-5114 Front Panel Indicators | |||||||
| Label | Function | For more information, refer to the NI High-Speed Digitizers Help. | |||||
| ACCESS | The ACCESS LED indicates the status of the PCI bus and the interface from the NI PXI-5114 to the controller. | ||||||
| ACTIVE | The ACTIVE LED indicates the status of the onboard acquisition hardware of the NI PXI-5114. | ||||||
| Dimensions and Weight |
©2010 National Instruments. All rights reserved. CompactRIO, CVI, FieldPoint, LabVIEW, Measurement Studio, National Instruments, National Instruments Alliance Partner, NI, and ni.com are trademarks of National Instruments. The mark LabWindows is used under a license from Microsoft Corporation. Windows is a registered trademark of Microsoft Corporation in the United States and other countries. Other product and company names listed are trademarks or trade names of their respective companies. A National Instruments Alliance Partner is a business entity independent from National Instruments and has no agency, partnership, or joint-venture relationship with National Instruments.