Last Revised: 2011-01-28 13:27:39.0
|
|
|
NI system assurance programs are designed to make it even easier for you to own an NI system. These programs include configuration and deployment services for your NI PXI, CompactRIO, or Compact FieldPoint system. The NI Basic System Assurance Program provides a simple integration test and ensures that your system is delivered completely assembled in one box. When you configure your system with the NI Standard System Assurance Program, you can select from available NI system driver sets and application development environments to create customized, reorderable software configurations. Your system arrives fully assembled and tested in one box with your software preinstalled. When you order your system with the standard program, you also receive system-specific documentation including a bill of materials, an integration test report, a recommended maintenance plan, and frequently asked question documents. Finally, the standard program reduces the total cost of owning an NI system by providing three years of warranty coverage and calibration service. Use the online product advisors at ni.com/advisor to find a system assurance program to meet your needs.
NI measurement hardware is calibrated to ensure measurement accuracy and verify that the device meets its published specifications. To ensure the ongoing accuracy of your measurement hardware, NI offers basic or detailed recalibration service that provides ongoing ISO 9001 audit compliance and confidence in your measurements. To learn more about NI calibration services or to locate a qualified service center near you, contact your local sales office or visit ni.com/calibration.
Get answers to your technical questions using the following National Instruments resources.
While you may never need your hardware repaired, NI understands that unexpected events may lead to necessary repairs. NI offers repair services performed by highly trained technicians who quickly return your device with the guarantee that it will perform to factory specifications. For more information, visit ni.com/repair.
The NI training and certification program delivers the fastest, most certain route to increased proficiency and productivity using NI software and hardware. Training builds the skills to more efficiently develop robust, maintainable applications, while certification validates your knowledge and ability.
NI offers options for extending the standard product warranty to meet the life-cycle requirements of your project. In addition, because NI understands that your requirements may change, the extended warranty is flexible in length and easily renewed. For more information, visit ni.com/warranty.
NI offers design-in consulting and product integration assistance if you need NI products for OEM applications. For information about special pricing and services for OEM customers, visit ni.com/oem.
Our Professional Services Team is comprised of NI applications engineers, NI Consulting Services, and a worldwide National Instruments Alliance Partner program of more than 700 independent consultants and integrators. Services range from start-up assistance to turnkey system integration. Visit ni.com/alliance.
| Output Characteristics | ||
|
Number of channels |
8 digital output channels | |
|
Output type |
Sourcing | |
|
Supply voltage range (Vsup) |
6 to 30 V | |
|
Output voltage |
Vsup – (I0 · R0) | |
|
Output impedance (R0) |
0.13 Ω max; 0.07 Ω typ | |
|
Continuous output current (I0) |
0.75 A max per channel | |
|
I/O protection |
||
|
Voltage |
30 V max | |
|
Reversed voltage |
None | |
|
Trip currents |
||
|
Minimum possible trip current |
6 A | |
|
Minimum guaranteed trip current |
13 A | |
|
Trip time |
10 μs at 13 A | |
| Current | Channel State | Module State |
|---|---|---|
| 0 to 1 A | Channel does not trip | Module is not damaged |
| 1 to 6 A | Channel does not trip | Module may be damaged |
| 6 to 13 A | Channel may trip | Module may be damaged |
| >13 A | Channel trips | Module is not damaged |
|
Output delay time (full load) |
100 μs max | |
| Power Requirements | ||
|
Current consumption from USB |
500 mA, max | |
|
Suspend mode |
2.5 mA, max | |
| Bus Interface | ||
|
USB specification |
USB 2.0 Hi-Speed | |
| Physical Characteristics | ||
|
Dimensions |
||
|
With combicon |
14.0 cm × 8.8 cm × 2.5 cm (5.52 in. × 3.47 in. × 1.00 in.) | |
|
With D-SUB |
12.6 cm × 8.8 cm × 2.5 cm (4.95 in. × 3.47 in. × 1.00 in.) | |
|
Screw terminal wiring |
12 to 24 AWG copper conductor wire with 10 mm (0.39 in.) of insulation stripped from the end | |
|
Torque for screw terminals |
0.5–0.6 N · m (4.4–5.3 lb · in.) | |
|
Weight |
||
|
With screw terminal |
Approx. 250 g (8.8 oz) | |
|
With D-SUB |
Approx. 245 g (8.6 oz) | |
| Safety | ||
| If you need to clean the module, wipe it with a dry towel. | ||
| Safety Voltages | ||
| NI USB-9472 with Screw Terminal | ||
|
Channel-to-COM |
30 V max | |
|
Isolation |
||
|
Channel-to-channel |
No isolation between channels | |
|
Channel-to-earth ground |
||
|
Withstand |
2,300 Vrms, 5 seconds max | |
|
Continuous |
250 Vrms, Measurement Category II | |
| Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet (e.g., 115 V for U.S. or 230 V for Europe). Examples of Measurement Category II are measurements performed on household appliances, portable tools, and similar products. | ||
| NI USB-9472 with D-SUB | ||
|
Channel-to-COM |
30 V max | |
|
Isolation |
||
|
Channel-to-channel |
No isolation between channels | |
|
Channel-to-earth ground |
||
|
Withstand |
1,000 Vrms, 5 seconds max | |
|
Continuous |
60 VDC, Measurement Category I | |
| Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS 1 voltage. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low-voltage sources, and electronics. | ||
| Safety Standards | ||
|
This product is designed to meet the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use:
Note For UL and other safety certifications, refer to the product label or the Online Product Certification section. |
||
| Hazardous Locations | ||
| This device is not certified for use in hazardous locations. | ||
| Environmental | ||
| The NI USB-9472 is intended for indoor use only. | ||
|
Operating temperature (IEC 60068-2-1, IEC 60068-2-2) |
0 to 60 °C | |
|
Storage temperature (IEC 60068-2-1, IEC 60068-2-2) |
– 40 to 85 °C | |
|
Operating humidity (IEC 60068-2-56) |
10 to 90% RH, noncondensing | |
|
Storage humidity (IEC 60068-2-56) |
5 to 95% RH, noncondensing | |
|
Maximum altitude |
2,000 m | |
|
Pollution Degree (IEC 60664) |
2 | |
| Electromagnetic Compatibility | ||
|
This product is designed to meet the requirements of the following standards of EMC for electrical equipment for measurement, control, and laboratory use:
Note For EMC compliance, operate this device with double-shielded cables. |
||
| CE Compliance | ||
|
This product meets the essential requirements of applicable European Directives, as amended for CE marking, as follows:
Note For the standards applied to assess the EMC of this product, refer to the Online Product Certification section. |
||
| Online Product Certification | ||
| Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, search by module number or product line, and click the appropriate link in the Certification column. | ||
| Environmental Management | ||
|
National Instruments is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial not only to the environment but also to NI customers. For additional environmental information, refer to the NI and the Environment Web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document. |
||
| Waste Electrical and Electronic Equipment (WEEE) | ||
|
EU Customers At the end of their life cycle, all products must be sent to a WEEE recycling center. For more information about WEEE recycling centers and National Instruments WEEE initiatives, visit ni.com/environment/weee.htm. ![]() |
||
![]() |
| NI USB-9472 Pin Assignments |
![]() |
| NI USB-9472 Terminal Assignments |