Customer SolutionsFunctional Testing of High Voltage X-Ray Tanks
Author(s):Roy Ignatius, Soliton Automation Private Limited
Industry:Life Science
Product:PXI/CompactPCI
The Challenge:Developing an automated test system to perform functional testing, of a high-voltage source tank for X-ray generators used in medical applications, involving high voltages.
The Solution:Creating a compact and cost effective PXI-based system using NI DMM, NI oscilloscope and data acquisition card, running a LabVIEW application.Introduction GE needed a test station that would perform all nine tests with minimal human intervention. ATE also needed to be able to communicate with the existing instruments that had computer connectivity through GPIB or RS 232 interface. We decided to build the ATE on a PXI platform to provide a compact yet cost effective replacement for the large manual test bay. The system was designed to restrict human intervention only to give the connections initially after which the system takes over and switches between the various instruments for the nine different tests. Further, the consolidated reports for GE’s Six-Sigma quality analysis were also generated from the ATE thereby significantly saving time for GE. The ATE was compactly designed using an eight slot PXI chassis with an embedded controller. A PXI switch switches between the different external instruments. PXI 6508 digital I/O module in combination with an electromechanical relay module was used to control switching that was needed for the various tests. The switching arrangement eliminated the need to change connections after every test. High voltage supplies were controlled by the relays through contactors. We used a PXI scope card to capture the waveforms. We used a PXI DMM to measure the high voltage by dropping it across a resistor and to measure resistances. A leakage tester that tests the system for leakage was controlled through the RS 232 interface. Through a GPIB interface, a high precision LCR meter from Wayne Kerr was used to measure HV capacitance, impedance, inductance and resonant frequencies. A programmable pulse generator was controlled through a GPIB interface for providing waveforms for different tests. The DAQ card measures the resultant voltages at different points and the results are compared with preset limits. Safety was of primary concern in the system due to the presence of high voltages up to 3,000V. We analyzed all failure conditions and provided additional safety features to take care of situations that might arise out of contactor failures, or power failures. Test sequencing was user selectable and the ATE would generate customized reports at the end of the test. The reports were required in Microsoft Excel format and using the Active X features of LabVIEW, the reports were directly created in the desired format in Excel. View the entire user solution in Adobe Acrobat PDF format. |