Test System Captures Heart Valve Sounds Using LabVIEW and DAQ Boards
Author(s):
Mark Ward - Structural Acoustics
Industry:
Biotechnology
Products:
LabVIEW,
The Challenge:
Developing a test system to test prosthetic heart valves.
The Solution:
Using LabVIEW to create a system that replicates the pumping action of the heart's left ventricle to load a prosthetic valve in either the aortic or mitral position, all of which contributes to building an acoustic diagnostic procedure for prosthetic heart valves. The procedure analyzes the clicking sound of a valve to determine the mechanical integrity of key components.
"The LabVIEW graphical programming language, which is highly efficient, becomes second nature very quickly."
| Structural Acoustics, a company that has been solving structure-born noise problems for 15 years, recently used LabVIEW for Windows in a unique electro-mechanical system for testing prosthetic heart valves. LabVIEW runs on the 486DX PC and acquires data while the system is pumping. The custom sensor interface takes care of signal conditioning. The sensor interface drives all of the sensors, amplifies their low- level signals, and provides an antialiasing filter before sending the voltage levels to the DAQ boards. Satisfied with the adjustments, we measure six signals during a typical test:
The acoustic signal is sampled at 100 ksamples per second (kS/s) by the first AT-MIO-16F. The other five signals are sampled at 40 kS/s by the second board. A digital trigger generated by the 386 PC paces the data acquisition. This PC executes a waveform editor written in C++ with which we can generate a digital trigger at any point during the heartbeat. Usually, there is one trigger per heartbeat placed about 20 ms before the valve closes. We are very pleased with our decision to use LabVIEW for Windows. It is a rich programming environment with excellent product support and documentation. The combination of LabVIEW for Windows and National Instruments hardware has performed very well for us. Mark Ward Structural Acoustics 3739 National Drive, Suite 100 Raleigh, NC 27612 Tel: (919) 783-8092 Fax: (919) 783-8352 E-mail: wardm@cs.unc.edu |
Related Case Studies
Caliper Brake Test System with an FPGA-based Servo Control System Using NI LabVIEWCreating a Test System for a New Generation Ultrasound Machine with LabVIEW and PXI
Developing a High-Voltage, Multi-Output Power Supply Test System with LabVIEW and PXI
Using National Instruments Software and Hardware to Develop an Automated Test System for Satellite Communication Equipment
On-Line Acoustical Quality Control System for Automobile Stereo Components Using LabVIEW and Neural Networks
|
|